Metadata | Start Date | End Date | Comment |
---|---|---|---|
CE07SHSM-SBD12-04-PCO2AA000 Method: recovered_host Stream: pco2a_a_dcl_instrument_air_recovered |
5/5/16, 3:30 PM | 9/27/16, 3:30 PM |
A vendor error in assembly pinched closed an internal tube carrying the surface air stream into the sample volume, resulting in an erroneous measurement. Users should not use the PCO2A air measurement for the Spring 2016 deployment (deployment 03), nor the associated sea-air pCO2 flux estimate.
Id: 1974
By: wingardc
|
CE07SHSM-SBD12-04-PCO2AA000 Method: recovered_host Stream: pco2a_a_dcl_instrument_air_recovered |
4/11/17, 1:30 PM | 10/6/17, 10:30 AM |
A vendor error in assembly pinched closed an internal tube carrying the surface air stream into the sample volume, resulting in an erroneous measurement. Users should not use the PCO2A air measurement for the Spring 2017 deployment (deployment 05), nor the associated sea-air pCO2 flux estimate.
Id: 1975
By: wingardc
|
CE07SHSM-SBD12-04-PCO2AA000 Method: recovered_host Stream: pco2a_a_dcl_instrument_air_recovered |
2/24/19, 3:30 AM | 2/24/19, 6:30 PM |
A HITL review of the data has found unusual, nonrealistic data during this time range. The data during this time range is marked as failed and users should avoid using this data in their analysis.
Id: 1976
By: wingardc
|
CE07SHSM-SBD12-04-PCO2AA000 Method: recovered_host Stream: pco2a_a_dcl_instrument_air_recovered |
3/4/19, 12:30 PM | 3/5/19, 3:30 AM |
A HITL review of the data has found unusual, nonrealistic data during this time range. The data during this time range is marked as failed and users should avoid using this data in their analysis.
Id: 1977
By: wingardc
|
CE07SHSM-SBD12-04-PCO2AA000 Method: recovered_host Stream: pco2a_a_dcl_instrument_air_recovered |
11/4/19, 11:30 AM | 11/5/19, 8:30 AM |
A HITL review of the data has found unusual, suspect data during this time range. The data during this time range is marked as suspect, and users should exercise caution using this data in their analysis.
Id: 1978
By: wingardc
|
Id | Parameter | Display Name | Standard Name | Unit | Fill Value | Precision | Identifier | Type | Level |
---|---|---|---|---|---|---|---|---|---|
PD7 | time | Time, UTC | time | seconds since 1900-01-01 | -9999999 | ||||
PD10 | port_timestamp | Port Timestamp, UTC | seconds since 1900-01-01 | -9999999 | |||||
PD11 | driver_timestamp | Driver Timestamp, UTC | seconds since 1900-01-01 | -9999999 | |||||
PD12 | internal_timestamp | Internal Timestamp, UTC | seconds since 1900-01-01 | -9999999 | |||||
PD16 | preferred_timestamp | Preferred Timestamp | 1 | empty | |||||
PD93 | date_time_string | Date and Time String | 1 | empty | |||||
PD863 | ingestion_timestamp | Ingestion Timestamp, UTC | seconds since 1900-01-01 | -9999 | |||||
PD992 | zero_a2d | Zero A/D | counts | -9999 | |||||
PD993 | current_a2d | Current A/D | counts | -9999 | |||||
PD994 | measured_air_co2 | CO2 Mole Fraction in Atmosphere | ppm | -9999999 | XCO2ATM_L0 | Science Data | L0 | ||
PD996 | avg_irga_temperature | Average IRGA Temperature | ºC | -9999999 | |||||
PD997 | humidity | Humidity | mbar | -9999999 | |||||
PD998 | humidity_temperature | Humidity Sensor Temperature | ºC | -9999999 | |||||
PD999 | gas_stream_pressure | Gas Stream Pressure | mbar | -9999 | PRESAIR_L0 | Unprocessed Data | L0 | ||
PD1000 | irga_detector_temperature | IRGA Detector Instrument | ºC | -9999999 | |||||
PD1001 | irga_source_temperature | IRGA Source Temperature | ºC | -9999999 | |||||
PD1002 | partial_pressure_co2_atm | pCO2 Atmosphere | µatm | -9999999 | PCO2ATM_L1 | Science Data | L1 | ||
PD2605 | dcl_controller_timestamp | DCL Controller Timestamp | 1 | empty | |||||
PD2805 | pco2_co2flux | CO2 Flux from Ocean to Atmosphere | mol m-2 s-1 | -9999999 | CO2FLUX_L2 | Science Data | L2 |