Data Stream Report

Instrument Name
Dissolved Oxygen
Reference Designator
CE09OSPM-WFP01-02-DOFSTK000
Method
telemetered
Stream
dofst_k_wfp_metadata
Stream Type
Engineering
Stream Content
Metadata
Uframe Route
Ingest.dofst-k-wfp_telemetered
Driver
Parser

Annotations

Metadata Start Date End Date Comment
CE09OSPM-WFP01-02-DOFSTK000
Method: telemetered
Stream: dofst_k_wfp_metadata
4/18/14, 8:00 AM 10/16/15, 7:00 AM

The SBE43 dissolved oxygen sensor (DOFST-K) was incorrectly assembled and stored resulting in the membrane drying out and tearing in one case. Clarification with the vendor on the proper assembly and storage techniques corrected the problem, however the data collected by the DOFST-K during deployments 1-3 is erroneous and should not be used. The data for this time period is excluded to avoid having it incorrectly used and disseminated.

Id: 1669 By: wingardc
Flag: fail Exclude: Yes