Metadata | Start Date | End Date | Comment |
---|---|---|---|
CP05MOAS-GL336-04-DOSTAM000 Method: telemetered Stream: dosta_abcdjm_glider_instrument |
9/8/18, 8:00 PM | 10/25/18, 7:59 PM |
The oxygen sensor on Glider CP05MOAS-GL336 began showing signs of failure around 9/9/18 and would stop sampling mid-profile. Upon further investigation, the sensor was creating errors internally on the glider and caused the mission to abort. To prevent further aborts, the sensor was shut down on 9/13/18. This glider was recovered on 10/25/18 and the sensor will be fixed prior to the next deployment.
Id: 1533
By: cdobson
|
Id | Parameter | Display Name | Standard Name | Unit | Fill Value | Precision | Identifier | Type | Level |
---|---|---|---|---|---|---|---|---|---|
PD7 | time | Time, UTC | time | seconds since 1900-01-01 | -9999999 | ||||
PD10 | port_timestamp | Port Timestamp, UTC | seconds since 1900-01-01 | -9999999 | |||||
PD11 | driver_timestamp | Driver Timestamp, UTC | seconds since 1900-01-01 | -9999999 | |||||
PD12 | internal_timestamp | Internal Timestamp, UTC | seconds since 1900-01-01 | -9999999 | |||||
PD16 | preferred_timestamp | Preferred Timestamp | 1 | empty | |||||
PD863 | ingestion_timestamp | Ingestion Timestamp, UTC | seconds since 1900-01-01 | -9999 | |||||
PD1401 | m_present_secs_into_mission | Elapsed Mission Time | s | ||||||
PD1402 | m_present_time | Time at the Start of the Cycle | seconds since 1970-01-01 | ||||||
PD1494 | sci_m_present_secs_into_mission | Elapsed Mission Time Based on Science Derived Start Times | s | ||||||
PD1495 | sci_m_present_time | Science Derived Time at the Start of the CycleC | seconds since 1970-01-01 | ||||||
PD1516 | sci_oxy4_oxygen | DO | µmol L-1 | DOCONCS_L1 | Science Data | L1 | |||
PD1518 | sci_oxy4_saturation | Dissolved Oxygen Saturation | percent | Science Data | |||||
PD2750 | sci_abs_oxygen | DO - Pressure Sal Corrected | moles_of_oxygen_per_unit_mass_in_sea_water | µmol kg-1 | DOXYGEN_L2 | Science Data | L2 |