Dep. | Preferred Method | Stream | DD | FD | SG | EG | Gaps | GD | TS | Rate (s) | Pressure Comp. | Time Order | Valid Data | Missing Data | Data Comp. | Missing Coords. | Review |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
1 | recovered_cspp | / | Complete | ||||||||||||||
3 | recovered_cspp | / | Complete | ||||||||||||||
4 | recovered_cspp | 88 | 44 | 0 | 0 | 13 | 47 | 79,040 | 80 / 3072 | 1 | 1 | Complete | |||||
5 | recovered_cspp | 32 | 18 | 0 | 9 | 3 | 6 | 35,279 | 80 / 1773 | 1 | 1 | Complete | |||||
6 | recovered_cspp | 5 | 4 | 0 | 1 | 0 | 0 | 14,878 | 80 / 2508 | 1 | 1 | Complete | |||||
7 | recovered_cspp | 105 | 44 | 0 | 40 | 4 | 22 | 118,475 | 80 / 78 | 1 | 1 | Complete | |||||
8 | recovered_cspp | 27 | 22 | 0 | 3 | 1 | 2 | 49,155 | 80 / 78 | 1 | 1 | Complete | |||||
9 | recovered_cspp | 20 | 19 | 0 | 1 | 0 | 0 | 49,102 | 80 / 78 | 1 | 1 | Complete | |||||
10 | recovered_cspp | 71 | 7 | 0 | 64 | 0 | 0 | 23,717 | 80 / 76 | 1 | 1 | Complete |
Deployment: | 1 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 |
---|---|---|---|---|---|---|---|---|---|
50% | 56% | 80% | 42% | 81% | 95% | 10% |
Deployment: | 1 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 |
---|---|---|---|---|---|---|---|---|---|
4 | 0.00 | ||||||||
5 | 0.19 | 0.00 | |||||||
6 | 0.42 | 0.40 | 0.00 | ||||||
7 | 0.49 | 0.37 | 0.25 | 0.00 | |||||
8 | 0.33 | 0.48 | 0.41 | 0.62 | 0.00 | ||||
9 | 0.67 | 0.51 | 0.45 | 0.21 | 0.83 | 0.00 | |||
10 | 0.00 |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
CE02SHSP-SP001-04-OPTAAJ000 |
8/17/16, 8:00 PM | 10/1/16, 8:00 PM |
OPTAA not plugged in, so no OPTAA data from this deployment, and the open cable and bulkhead may require repair.
Id: 884
By: michaesm
|
CE02SHSP |
3/8/17, 7:00 PM | 4/17/17, 8:00 PM |
The main batteries on CE02SHSP-00005 drained unexpectedly. CSPPs are deployed at 33.4 V. Voltage drops roughly linearly with each profile. Batteries were at 31 V at the beginning of this week. Yesterday (2017-03-09), it reported via acoustic modem that it was in a LOW POWER state, which it does automatically at 28 V to protect the main batteries. In the past, similar drains have been due to faulty connections between the main batteries and the profilers at the battery adapter plates. This has been the CSPP's main failure behavior.
Id: 882
By: michaesm
|
CE02SHSP |
4/24/17, 8:00 PM | 10/4/17, 8:00 PM |
CE02SHSP-00006 reported that it had surfaced unexpectedly April 25 at 5:10 pm local time. Within a few hours it was determined that it was still tethered to its anchor. It's beacon wasn't reporting either. The profiler and its anchor were recovered the following day. Data were sent to the profiler vendor. By April 28 we confirmed there was a firmware bug that had never been seen before. While moving between two positions, the control canister thought the profiler was 20 cm below its destination, but the profiler's winch thought the profiler was 20 cm above its destination, so the profiler was winched up to the surface, instead of retreating 20 cm to its parking depth. The bug will be corrected in the next firmware release. It also turns out that the beacon was programmed improperly, so it had run out of batteries. We switched to a new beacon model this year and confirmed with the vendor which settings to use. Id: 883 By: michaesm |
CE02SHSP |
5/29/18, 8:00 PM | 8/1/18, 4:30 PM |
Profiler was profiling twice per day, but it has not surfaced as planned since May 30 ~1am PDT. It is deployed too far from CE02SHSM for us to have good acoustic communications (1150m vs. the planned 150-400m), so we cannot debug it remotely. It will need to be recovered, but it is not at the surface or adrift, so it does not need immediate recovery. Id: 1458 By: michaesm |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
CE02SHSP-SP001 |
|||
CE02SHSP-SP001-04-OPTAAJ000 |
According to Roesler and Barnard (Optical proxy for phytoplankton biomass in the absence of photophysiology: Rethinking the absorption line height, Methods in Oceanography, 2013), “absorption meters are highly prone to biofouling, particularly biofilms which not only attenuate the collimated beam but also impact the scattering properties of the optical surfaces and tubes. In productive coastal waters biofouling can have significant impacts (i.e. 10% of the signal) within one to two weeks”. By Lori Garzio, on 7/1/19 |
||
CE02SHSP-SP001-04-OPTAAJ000 |
The file coordinate pressure_depth is an array of all zeros for every deployment. The variable in the file called int_ctd_pressure has what appears to be valid pressure data. The coordinate variable pressure_depth should contain valid pressure data. By Lori Garzio, on 7/1/19 |
||
CE02SHSP-SP001-04-OPTAAJ000 Deployment: 1 |
There are no data available for download from deployment 1. According to the ingest csv, the raw data files are available. These files should be ingested or the dataset should be annotated. By Lori Garzio, on 7/1/19 |
||
CE02SHSP-SP001-04-OPTAAJ000 Deployment: 4 |
60% of beam_attenuation and 63% of optical_absorption data for this deployment are nans, fill values, outside of global ranges, or outliers (+/- 5 SD). By Lori Garzio, on 7/1/19 |
||
CE02SHSP-SP001-04-OPTAAJ000 Deployment: 7 |
87% of beam_attenuation and optical_absorption data for this deployment are nans, fill values, outside of global ranges, or outliers (+/- 5 SD). By Lori Garzio, on 7/1/19 |