Seawater pH

Reference Designator
CE07SHSM-RID26-06-PHSEND000
Review Status
Review Complete
Note
Depth
7m
Class
PHSEN (Seawater pH)
Make / Model
Sunburst / SAMI-pH

Dataset Reviews Last processed: 5/15/19, 2:05 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 recovered_inst 162 162 0 0 0 0 3,862 7 / 1 Complete
2 recovered_inst 214 213 0 1 0 0 7,602 7 / 2 1 Complete
3 recovered_inst 146 146 0 0 0 0 4,595 7 / 1 Complete
4 recovered_inst 206 144 0 62 0 0 3,340 7 / 1 Complete
5 recovered_inst 179 178 0 0 1 2 3,943 7 / 1 Complete
6 recovered_inst 178 140 0 38 0 0 3,524 7 / 1 Complete
7 recovered_host 180 171 0 0 2 9 3,964 7 / 1 Complete
7 recovered_inst / Complete
Data Ranges Review Images

Test Notes

  1. missing: ['pressure']
  2. fail: data found in another stream (gaps: [1] days: [2])

Data Coverage

Deployment: 1234567
95%
100%100%100%70%99%79%

Lat/Lon Differences (km)

Deployment: 1234567
1 0.00
2 0.280.00
3 0.440.220.00
4 0.020.290.460.00
5 0.250.170.210.270.00
6 0.180.460.610.170.400.00
7 0.270.100.170.290.080.440.00

System Annotations

Metadata Start Date End Date Comment
CE07SHSM
12/1/17, 9:35 AM 3/25/18, 3:09 PM

Data loss due to a power system failure. The mooring stopped receiving power from photo voltaic panel 1 of 4 on Nov-2. By Nov-20, the mooring stopped receiving power from two more solar panels (PV2 & PV4) and one of two wind turbines (WT2). This indicates the Power System Controller (PSC) they connect to is failing. Mooring components will be shut down as the mooring loses the ability to recharge its batteries.

Id: 1237 By: crisien
Flag: not_operational Exclude: No

CE07SHSM-RID26-06-PHSEND000
8/4/19, 4:35 AM 8/5/19, 9:35 PM

Data indicates the sensor has either fouled or the intake tubing has become clogged.

Id: 1699 By: wingardc
Flag: suspect Exclude: No

CE07SHSM
Method: telemetered
5/13/20, 6:40 PM

Due to a bug in the mooring telemetry system, data during this time range was not telemetered to shore. It was recorded on the mooring and instruments and will be available in the recovered data streams.

Id: 2185 By: wingardc
Flag: not_available Exclude: No

Review Notes

Metadata Start Date End Date Comment
CE07SHSM-RID26-06-PHSEND000

recovered instrument data from deployment 7 are missing

By Leila Belabbassi, on 5/23/19

CE07SHSM-RID26-06-PHSEND000

a large gap at the end of the deployment is not annotated in the system.

By Leila Belabbassi, on 5/24/19

CE07SHSM-RID26-06-PHSEND000

data gaps are not annotated in the system (deployments 5 and 7).

By Leila Belabbassi, on 5/24/19

CE07SHSM-RID26-06-PHSEND000
3/1/16, 12:00 AM 10/31/16, 12:00 AM

suspect data

By Leila Belabbassi, on 5/24/19

CE07SHSM-RID26-06-PHSEND000
4/1/17, 12:00 AM 9/30/18, 12:00 AM

suspect data

By Leila Belabbassi, on 5/24/19

CE07SHSM-RID26-06-PHSEND000
Deployment: 1
4/15/15, 12:00 AM 9/23/15, 12:00 AM

suspect data

By Leila Belabbassi, on 5/24/19

CE07SHSM-RID26-06-PHSEND000
Deployment: 2

recovered_inst data are not complete. Data are found in the recovered_host stream.

By Leila Belabbassi, on 5/24/19

New Note