3-Wavelength Fluorometer

Reference Designator
CE07SHSM-RID27-02-FLORTD000
Review Status
Review Complete
Note
Depth
7m
Class
FLORT (3-Wavelength Fluorometer)
Make / Model
WET Labs / ECO Triplet-w

Dataset Reviews Last processed: 6/27/19, 1:19 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 recovered_host flort_sample 162 162 0 0 0 0 2,296,963 1 7 / Complete
2 recovered_host flort_sample 214 161 0 0 2 54 2,248,031 1 7 / Complete
3 recovered_host flort_sample 146 146 0 0 0 0 2,071,514 1 7 / Complete
4 recovered_host flort_sample 206 96 0 110 0 0 1,350,870 1 7 / Complete
5 recovered_host flort_sample 179 120 0 15 5 46 1,377,923 1 7 / Complete
6 recovered_host flort_sample 178 106 0 71 0 0 1,370,118 1 7 / Complete
7 recovered_host flort_sample 180 171 0 0 2 9 1,667,663 1 7 / Complete
Data Ranges Review Images

Test Notes

All good!

Data Coverage

Deployment: 1234567
flort_sample 100%75%100%47%67%60%95%

Lat/Lon Differences (km)

Deployment: 1234567
1 0.00
2 0.280.00
3 0.440.220.00
4 0.020.290.460.00
5 0.250.170.210.270.00
6 0.180.460.610.170.400.00
7 0.270.100.170.290.080.440.00

Example Composite Data Plot

System Annotations

Metadata Start Date End Date Comment
CE07SHSM
12/1/17, 9:35 AM 3/25/18, 3:09 PM

Data loss due to a power system failure. The mooring stopped receiving power from photo voltaic panel 1 of 4 on Nov-2. By Nov-20, the mooring stopped receiving power from two more solar panels (PV2 & PV4) and one of two wind turbines (WT2). This indicates the Power System Controller (PSC) they connect to is failing. Mooring components will be shut down as the mooring loses the ability to recharge its batteries.

Id: 1237 By: crisien
Flag: not_operational Exclude: No

CE07SHSM
Method: telemetered
5/13/20, 6:40 PM

Due to a bug in the mooring telemetry system, data during this time range was not telemetered to shore. It was recorded on the mooring and instruments and will be available in the recovered data streams.

Id: 2185 By: wingardc
Flag: not_available Exclude: No

Review Notes

Metadata Start Date End Date Comment
CE07SHSM-RID27-02-FLORTD000

pressure_depth is an array of fill values for every deployment. This variable should contain valid pressure data.

By Lori Garzio, on 7/2/19

CE07SHSM-RID27-02-FLORTD000
Deployment: 2
3/20/16, 12:00 AM 5/10/15, 12:00 AM

Missing data should be annotated

By Lori Garzio, on 7/2/19

CE07SHSM-RID27-02-FLORTD000
Deployment: 2
5/10/16, 12:00 AM 5/13/16, 10:15 PM

Suspect data at the end of the deployment.

By Lori Garzio, on 7/2/19

CE07SHSM-RID27-02-FLORTD000
Deployment: 4
12/24/16, 2:18 PM 4/13/17, 11:22 PM

Missing data at the end of the deployment should be annotated.

By Lori Garzio, on 7/2/19

CE07SHSM-RID27-02-FLORTD000
Deployment: 5

A substantial amount of CDOM values are negative

By Lori Garzio, on 7/2/19

CE07SHSM-RID27-02-FLORTD000
Deployment: 5

Data gaps should be annotated

By Lori Garzio, on 7/2/19

CE07SHSM-RID27-02-FLORTD000
Deployment: 5
8/15/17, 12:00 AM 10/6/17, 2:26 PM

Suspect data at the end of the deployment. total_volume_scattering_coefficient flatlined at the upper detection limit

By Lori Garzio, on 7/2/19

CE07SHSM-RID27-02-FLORTD000
Deployment: 6
12/29/17, 12:00 AM 1/20/18, 12:00 AM

Suspect data at the end of the deployment. total_volume_scattering_coefficient flatlined at the upper detection limit

By Lori Garzio, on 7/2/19

New Note