pCO2 Air-Sea

Reference Designator
CE07SHSM-SBD12-04-PCO2AA000
Review Status
Review Complete
Note
Depth
0m
Class
PCO2A (pCO2 Air-Sea)
Make / Model
Pro-Oceanus / pCO2-pro

Dataset Reviews Last processed: 7/23/19, 7:28 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 recovered_host 162 162 0 0 0 0 34,758 1 / 1 1 Complete
1 recovered_host 162 162 0 0 0 0 34,758 1 1 / 1 1 Complete
2 recovered_host 214 214 0 0 0 0 46,053 1 / 1 1 Complete
2 recovered_host 214 214 0 0 0 0 46,053 1 1 / 1 1 Complete
3 recovered_host 146 146 0 0 0 0 31,320 1 / 1 1 Complete
3 recovered_host 146 146 0 0 0 0 31,319 1 1 / 1 1 Complete
4 recovered_host 206 28 54 123 0 0 6,048 1 / 2 1 1 Complete
4 recovered_host 206 28 54 123 0 0 6,048 1 1 / 2 1 1 Complete
5 telemetered 179 95 0 16 7 72 18,198 1 / 1 1 Complete
5 telemetered 179 95 0 16 7 72 18,197 1 1 / 1 1 Complete
6 recovered_host 178 28 68 80 0 0 6,048 1 / 2 1 1 Complete
6 recovered_host 178 28 68 80 0 0 6,048 1 1 / 2 1 1 Complete
7 recovered_host 180 128 0 43 1 8 25,965 1 / 1 1 Complete
7 recovered_host 180 128 0 43 1 8 25,960 1 1 / 1 1 Complete
Data Ranges Review Images

Test Notes

  1. no other streams for comparison
  2. pass unique test - ascending not tested

Data Coverage

Deployment: 1234567
53%
100%100%100%14%16%71%
53%
100%100%100%14%16%71%

Lat/Lon Differences (km)

Deployment: 1234567
1 0.00
2 0.280.00
3 0.440.220.00
4 0.020.290.460.00
5 0.250.170.210.270.00
6 0.180.460.610.170.400.00
7 0.270.100.170.290.080.440.00

Example Composite Data Plot

System Annotations

Metadata Start Date End Date Comment
CE07SHSM-SBD12-04-PCO2AA000
Method: telemetered
Stream: pco2a_a_dcl_instrument_air
4/5/15, 1:30 PM 9/23/15, 11:30 AM

PCO2A air measurement for the Spring 2015 deployment is considered suspect. Recent investigations have indicated a vendor error in assembly may have pinched closed an internal tube carrying the surface air stream into the sample volume resulting in an erroneous measurement.

Id: 1968 By: wingardc
Flag: suspect Exclude: No

CE07SHSM-SBD12-04-PCO2AA000
Method: telemetered
Stream: pco2a_a_dcl_instrument_air
5/5/16, 3:30 PM 9/27/16, 3:30 PM

A vendor error in assembly pinched closed an internal tube carrying the surface air stream into the sample volume, resulting in an erroneous measurement. Users should not use the PCO2A air measurement for the Spring 2016 deployment (deployment 03), nor the associated sea-air pCO2 flux estimate.

Id: 1969 By: wingardc
Flag: fail Exclude: No

CE07SHSM-SBD12-04-PCO2AA000
Method: recovered_host
Stream: pco2a_a_dcl_instrument_air_recovered
5/5/16, 3:30 PM 9/27/16, 3:30 PM

A vendor error in assembly pinched closed an internal tube carrying the surface air stream into the sample volume, resulting in an erroneous measurement. Users should not use the PCO2A air measurement for the Spring 2016 deployment (deployment 03), nor the associated sea-air pCO2 flux estimate.

Id: 1974 By: wingardc
Flag: fail Exclude: No

CE07SHSM-SBD12-04-PCO2AA000
Method: telemetered
Stream: pco2a_a_dcl_instrument_water
5/5/16, 3:30 PM 9/27/16, 3:30 PM

A vendor error in assembly pinched closed an internal tube carrying the surface air stream into the sample volume, resulting in an erroneous measurement. Users should not use the PCO2A air measurement for the Spring 2016 deployment (deployment 03), nor the associated sea-air pCO2 flux estimate.

Id: 1992 By: wingardc
Flag: fail Exclude: No

CE07SHSM-SBD12-04-PCO2AA000
Method: recovered_host
Stream: pco2a_a_dcl_instrument_water_recovered
5/5/16, 3:30 PM 9/27/16, 3:30 PM

A vendor error in assembly pinched closed an internal tube carrying the surface air stream into the sample volume, resulting in an erroneous measurement. Users should not use the PCO2A air measurement for the Spring 2016 deployment (deployment 03), nor the associated sea-air pCO2 flux estimate.

Id: 1994 By: wingardc
Flag: fail Exclude: No

CE07SHSM-SBD12-04-PCO2AA000
Method: telemetered
Stream: pco2a_a_dcl_instrument_air
4/11/17, 1:30 PM 10/6/17, 10:30 AM

A vendor error in assembly pinched closed an internal tube carrying the surface air stream into the sample volume, resulting in an erroneous measurement. Users should not use the PCO2A air measurement for the Spring 2017 deployment (deployment 05), nor the associated sea-air pCO2 flux estimate.

Id: 1970 By: wingardc
Flag: fail Exclude: No

CE07SHSM-SBD12-04-PCO2AA000
Method: recovered_host
Stream: pco2a_a_dcl_instrument_air_recovered
4/11/17, 1:30 PM 10/6/17, 10:30 AM

A vendor error in assembly pinched closed an internal tube carrying the surface air stream into the sample volume, resulting in an erroneous measurement. Users should not use the PCO2A air measurement for the Spring 2017 deployment (deployment 05), nor the associated sea-air pCO2 flux estimate.

Id: 1975 By: wingardc
Flag: fail Exclude: No

CE07SHSM-SBD12-04-PCO2AA000
Method: telemetered
Stream: pco2a_a_dcl_instrument_water
4/11/17, 1:30 PM 10/6/17, 10:30 AM

A vendor error in assembly pinched closed an internal tube carrying the surface air stream into the sample volume, resulting in an erroneous measurement. Users should not use the PCO2A air measurement for the Spring 2017 deployment (deployment 05), nor the associated sea-air pCO2 flux estimate.

Id: 1993 By: wingardc
Flag: fail Exclude: No

CE07SHSM-SBD12-04-PCO2AA000
Method: recovered_host
Stream: pco2a_a_dcl_instrument_water_recovered
4/11/17, 1:30 PM 10/6/17, 10:30 AM

A vendor error in assembly pinched closed an internal tube carrying the surface air stream into the sample volume, resulting in an erroneous measurement. Users should not use the PCO2A air measurement for the Spring 2017 deployment (deployment 05), nor the associated sea-air pCO2 flux estimate.

Id: 1995 By: wingardc
Flag: fail Exclude: No

CE07SHSM
12/1/17, 9:35 AM 3/25/18, 3:09 PM

Data loss due to a power system failure. The mooring stopped receiving power from photo voltaic panel 1 of 4 on Nov-2. By Nov-20, the mooring stopped receiving power from two more solar panels (PV2 & PV4) and one of two wind turbines (WT2). This indicates the Power System Controller (PSC) they connect to is failing. Mooring components will be shut down as the mooring loses the ability to recharge its batteries.

Id: 1237 By: crisien
Flag: not_operational Exclude: No

CE07SHSM-SBD12-04-PCO2AA000
Method: telemetered
Stream: pco2a_a_dcl_instrument_air
2/24/19, 3:30 AM 2/24/19, 6:30 PM

A HITL review of the data has found unusual, nonrealistic data during this time range. The data during this time range is marked as failed and users should avoid using this data in their analysis.

Id: 1971 By: wingardc
Flag: fail Exclude: No

CE07SHSM-SBD12-04-PCO2AA000
Method: recovered_host
Stream: pco2a_a_dcl_instrument_air_recovered
2/24/19, 3:30 AM 2/24/19, 6:30 PM

A HITL review of the data has found unusual, nonrealistic data during this time range. The data during this time range is marked as failed and users should avoid using this data in their analysis.

Id: 1976 By: wingardc
Flag: fail Exclude: No

CE07SHSM-SBD12-04-PCO2AA000
Method: telemetered
Stream: pco2a_a_dcl_instrument_air
3/4/19, 12:30 PM 3/5/19, 3:30 AM

A HITL review of the data has found unusual, nonrealistic data during this time range. The data during this time range is marked as failed and users should avoid using this data in their analysis.

Id: 1972 By: wingardc
Flag: fail Exclude: No

CE07SHSM-SBD12-04-PCO2AA000
Method: recovered_host
Stream: pco2a_a_dcl_instrument_air_recovered
3/4/19, 12:30 PM 3/5/19, 3:30 AM

A HITL review of the data has found unusual, nonrealistic data during this time range. The data during this time range is marked as failed and users should avoid using this data in their analysis.

Id: 1977 By: wingardc
Flag: fail Exclude: No

CE07SHSM-SBD12-04-PCO2AA000
Method: telemetered
4/24/19, 11:12 AM 10/12/19, 7:51 PM

Due to a firmware bug, the sensor zeroing function for correcting the baseline drift over time is not currently saving the updated reference value used to compute the raw CO2 value outputs during the measurements. The sensors are zeroing at the end of the sample cycle when scheduled, but the firmware is not storing the new value. This results in the original zero reference value being used for all measurements and an observable and consistent negative drift in both the air and water CO2 values. The vendor has indicated that the data may be correctable post-recovery, however at this time no such correction algorithm has been provided.

Id: 1687 By: wingardc
Flag: fail Exclude: Yes

CE07SHSM-SBD12-04-PCO2AA000
Method: telemetered
Stream: pco2a_a_dcl_instrument_air
11/4/19, 11:30 AM 11/5/19, 8:30 AM

A HITL review of the data has found unusual, suspect data during this time range. The data during this time range is marked as suspect, and users should exercise caution using this data in their analysis.

Id: 1973 By: wingardc
Flag: suspect Exclude: No

CE07SHSM-SBD12-04-PCO2AA000
Method: recovered_host
Stream: pco2a_a_dcl_instrument_air_recovered
11/4/19, 11:30 AM 11/5/19, 8:30 AM

A HITL review of the data has found unusual, suspect data during this time range. The data during this time range is marked as suspect, and users should exercise caution using this data in their analysis.

Id: 1978 By: wingardc
Flag: suspect Exclude: No

CE07SHSM-SBD12-04-PCO2AA000
Method: telemetered
Stream: pco2a_a_dcl_instrument_water
11/4/19, 11:30 AM 11/5/19, 8:30 AM

A HITL review of the data has found unusual, suspect data during this time range. The data during this time range is marked as suspect, and users should exercise caution using this data in their analysis.

Id: 1979 By: wingardc
Flag: suspect Exclude: No

CE07SHSM-SBD12-04-PCO2AA000
Method: recovered_host
Stream: pco2a_a_dcl_instrument_water_recovered
11/4/19, 11:30 AM 11/5/19, 8:30 AM

A HITL review of the data has found unusual, suspect data during this time range. The data during this time range is marked as suspect, and users should exercise caution using this data in their analysis.

Id: 1980 By: wingardc
Flag: suspect Exclude: No

CE07SHSM-SBD12-04-PCO2AA000
Method: telemetered
11/10/19, 2:30 PM 11/12/19, 3:30 PM

The data logger port providing power to the instrument showed an over-current error and disabled the instrument. The port had to be power cycled in order to clear the error and restore operations.

Id: 2054 By: wingardc
Flag: not_operational Exclude: No

CE07SHSM-SBD12-04-PCO2AA000
Method: telemetered
3/7/20, 2:30 PM 3/10/20, 4:30 PM

The instrument suffered a firmware crash and had to be power cycled in order to restore operations. We are in consultation with the vendor to determine the cause of this failure.

Id: 2057 By: wingardc
Flag: not_operational Exclude: No

CE07SHSM-SBD12-04-PCO2AA000
Method: telemetered
3/30/20, 8:30 PM 3/31/20, 7:30 PM

The instrument suffered a firmware crash and had to be power cycled in order to restore operations. We are in consultation with the vendor to determine the cause of this failure.

Id: 2058 By: wingardc
Flag: not_operational Exclude: No

CE07SHSM-SBD12-04-PCO2AA000
Method: telemetered
4/5/20, 6:30 PM 4/8/20, 7:30 PM

The instrument suffered a firmware crash and had to be power cycled in order to restore operations. We are in consultation with the vendor to determine the cause of this failure.

Id: 2059 By: wingardc
Flag: not_operational Exclude: No

CE07SHSM
Method: telemetered
5/13/20, 6:40 PM

Due to a bug in the mooring telemetry system, data during this time range was not telemetered to shore. It was recorded on the mooring and instruments and will be available in the recovered data streams.

Id: 2185 By: wingardc
Flag: not_available Exclude: No

Review Notes

Metadata Start Date End Date Comment
CE07SHSM-SBD12-04-PCO2AA000
Deployment: 1
4/15/15, 5:15 PM 9/23/15, 3:08 PM

Suspect data: measured_air_co2 values decrease to 340 ppm by the end of the deployment

By Lori Garzio, on 7/31/19

CE07SHSM-SBD12-04-PCO2AA000
Deployment: 2
3/15/16, 12:00 AM 5/13/16, 10:15 PM

Suspect data - measured_water_co2 values drop off towards the end of the deployment.

By Lori Garzio, on 7/31/19

CE07SHSM-SBD12-04-PCO2AA000
Deployment: 3
5/5/16, 7:08 PM 9/27/16, 7:47 PM

measured_air_co2 values are too low for the majority of the deployment.

By Lori Garzio, on 7/31/19

CE07SHSM-SBD12-04-PCO2AA000
Deployment: 4

pco2_co2flux is all fill values for deployments 4 and 6.

By Lori Garzio, on 7/31/19

CE07SHSM-SBD12-04-PCO2AA000
Deployment: 4

Missing data for deployment 4 should be annotated.

By Lori Garzio, on 7/31/19

CE07SHSM-SBD12-04-PCO2AA000
Deployment: 5

recovered_host data are not available for download for this deployment. According to the ingest csv, the recovered_host raw data files are missing. These files should be provided and ingested.

By Lori Garzio, on 7/31/19

CE07SHSM-SBD12-04-PCO2AA000
Deployment: 5
4/11/17, 5:51 PM 10/6/17, 2:26 PM

measured_air_co2 values are too low.

By Lori Garzio, on 7/31/19

New Note