3-Wavelength Fluorometer

Reference Designator
CP01CNPM-WFP01-04-FLORTK000
Review Status
Review Complete
Note
Depth
15 to 70m
Class
FLORT (3-Wavelength Fluorometer)
Make / Model
WET Labs / ECO Triplet

Dataset Reviews Last processed: 12/18/18, 1:52 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 recovered_wfp flort_sample 158 158 0 0 0 0 163,317 6 104 / Complete
Data Ranges Review Images

Test Notes

All good!

Data Coverage

Deployment: 1
flort_sample 100%

Lat/Lon Differences (km)

Deployment: 1
1 0.00

System Annotations

Metadata Start Date End Date Comment
CP01CNPM-WFP01
12/31/18, 7:00 PM 2/6/19, 7:00 PM

Deployment 2: The WFP slid up the wire between profiles on various dates including 2019-01-01, 2019-01-29, and from 2019-02-05 through 2019-02-07.

Id: 1623 By: cdobson

CP01CNPM-WFP01
2/10/19, 7:00 PM 2/17/19, 7:00 PM

Deployment 2: The WFP becamestuck at 25m from 2019-02-11 through 2019-02-18.

Id: 1622 By: cdobson

CP01CNPM-WFP01
2/20/19, 7:00 PM 4/4/19, 8:07 AM

Deployment 2: Telemetry of data from the WFP ceased on 2019-02-21.

Id: 1621 By: cdobson

CP01CNPM-WFP01
12/2/19, 7:00 PM 6/14/20, 8:23 AM

Deployment 3: Starting 12/03/2019 WFP01 was unable to complete full range descending profiles.

Id: 1842 By: cdobson
Flag: not_operational Exclude: No

CP01CNPM-WFP01
12/31/19, 7:00 PM

Deployment 3: Due to a clock rollover issue, all WFPs have ceased sampling on 1/1/2020. No WFP data are being collected after this date. A resolution is being researched by the vendor.

Id: 1879 By: cdobson

Review Notes

Metadata Start Date End Date Comment
CP01CNPM-WFP01-04-FLORTK000

Comparison with the shipboard CTD cast was done for D0001 and R0001. For D0001, Chl-a recorded by the profiler was generally comparable to the CTD cast, with the exception of the chl-a maximum (~20 - 40 dbar) where the profiler recorded higher chl-a values. For R0001 the chl-a recorded from the profiler was higher. Further analysis using the shipboard bottle data should be conducted. This is outside the scope of this project.

By Lori Garzio, on 3/12/19

CP01CNPM

system missing global ranges for all parameters because the platform is not listed in the data_qc_global_range_values.csv

By Leila Belabbassi, on 3/15/19

CP01CNPM-WFP01-04-FLORTK000
Deployment: 1

total_volume_scattering_coefficient_m-1 sr-1 and optical_backscatter_m-1 show an event in mid March 2018 that need to be investigated further.

By Leila Belabbassi, on 3/18/19

CP01CNPM-WFP01-04-FLORTK000
Deployment: 1

pressure record in the data file are all NaNs. lgarzio edit: The pressure coordinate (pressure_depth) is an array of all fill values. This variable should contain valid pressure data. There is another variable in the files called in_ctd_pressure that contains what appears to be valid pressure data. This is the case for every FLORTK on every WFP. https://opendap.oceanobservatories.org/thredds/dodsC/ooi/lgarzio@marine.rutgers.edu/20181218T135235-CP01CNPM-WFP01-04-FLORTK000-recovered_wfp-flort_sample/deployment0001_CP01CNPM-WFP01-04-FLORTK000-recovered_wfp-flort_sample_20171106T190203-20180412T120904.nc.ascii?pressure_depth[0:1:1000],int_ctd_pressure[0:1:1000]

By Leila Belabbassi, on 3/6/19

New Note