Photosynthetically Available Radiation

Reference Designator
CP01CNPM-WFP01-05-PARADK000
Review Status
Review Complete
Note
Depth
15 to 70m
Class
PARAD (Photosynthetically Available Radiation)
Make / Model
Biospherical Instruments / QSP-2200

Dataset Reviews Last processed: 12/18/18, 1:53 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 recovered_wfp 158 158 0 0 0 0 163,317 6 104 / 116 Complete
Data Ranges Review Images

Test Notes

All good!

Data Coverage

Deployment: 1
100%

Lat/Lon Differences (km)

Deployment: 1
1 0.00

System Annotations

Metadata Start Date End Date Comment
CP01CNPM-WFP01
12/31/18, 7:00 PM 2/6/19, 7:00 PM

Deployment 2: The WFP slid up the wire between profiles on various dates including 2019-01-01, 2019-01-29, and from 2019-02-05 through 2019-02-07.

Id: 1623 By: cdobson

CP01CNPM-WFP01
2/10/19, 7:00 PM 2/17/19, 7:00 PM

Deployment 2: The WFP becamestuck at 25m from 2019-02-11 through 2019-02-18.

Id: 1622 By: cdobson

CP01CNPM-WFP01
2/20/19, 7:00 PM 4/4/19, 8:07 AM

Deployment 2: Telemetry of data from the WFP ceased on 2019-02-21.

Id: 1621 By: cdobson

CP01CNPM-WFP01
12/2/19, 7:00 PM 6/14/20, 8:23 AM

Deployment 3: Starting 12/03/2019 WFP01 was unable to complete full range descending profiles.

Id: 1842 By: cdobson
Flag: not_operational Exclude: No

CP01CNPM-WFP01
12/31/19, 7:00 PM

Deployment 3: Due to a clock rollover issue, all WFPs have ceased sampling on 1/1/2020. No WFP data are being collected after this date. A resolution is being researched by the vendor.

Id: 1879 By: cdobson

Review Notes

Metadata Start Date End Date Comment
CP01CNPM-WFP01-05-PARADK000

The depth recorded in the deployment sheet is 12 m shallower than the maximum depth the profiler is measuring. Is that by design?

By Leila Belabbassi, on 3/15/19

CP01CNPM-WFP01-05-PARADK000

The PAR measurements at depth greater than 100 dbar are too high [ > 5 - 30 µmol m-1 s-1] resembling the surface data.
The plot show data filtered out of nans, extreme values (1e7), fill values and values outside global ranges [ 0 - 2500] and 5 STD.

By Leila Belabbassi, on 3/15/19

CP01CNPM

system missing global ranges for all parameters because the platform is not listed in the data_qc_global_range_values.csv

By Leila Belabbassi, on 3/15/19