Dep. | Preferred Method | Stream | DD | FD | SG | EG | Gaps | GD | TS | Rate (s) | Pressure Comp. | Time Order | Valid Data | Missing Data | Data Comp. | Missing Coords. | Review |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
1 | recovered_wfp | 151 | 151 | 0 | 0 | 0 | 0 | 1,005,405 | 1 | 70 / 71 | Complete |
All good!
Deployment: | 1 |
---|---|
100% |
Deployment: | 1 |
---|---|
1 | 0.00 |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
CP03ISPM-WFP01 |
12/31/19, 7:00 PM | 6/13/20, 3:24 PM |
Deployment 3: Due to a clock rollover issue, all WFPs have ceased sampling on 1/1/2020. No WFP data are being collected after this date. A resolution is being researched by the vendor. Id: 1880 By: cdobson |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
CP03ISPM-WFP01-03-CTDPFK000 |
Comparisons with the shipboard CTD casts were done for D0001 and R0001. The profiles do not match the CTD cast for D0001, and more closely align with the CTD cast for R0001. This pattern is consistent with the profiler's DOFST (data do not match the shipboard CTD cast at deployment and match at recovery). Comparison to the shipboard bottle data is outside the scope of this project. By Lori Garzio, on 3/12/19 |