Bio-acoustic Sonar (Coastal)

Reference Designator
CP03ISSM-MFD37-07-ZPLSCC000
Review Status
Will Not Review
Note
Not Delivered Via uFrame
Depth
91.5m
Class
ZPLSC (Bio-acoustic Sonar (Coastal))
Make / Model
ASL Environmental Sciences / AZFP

Dataset Reviews Last processed:

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
3 telemetered zplsc_c_instrument / Todo
4 telemetered zplsc_c_instrument / Todo
5 telemetered zplsc_c_instrument / Todo
6 telemetered zplsc_c_instrument / Todo
7 telemetered zplsc_c_instrument / Todo
Data Ranges Review Images

Test Notes

All good!

System Annotations

Metadata Start Date End Date Comment
CP03ISSM-MFD37-07-ZPLSCC000
Stream: zplsc_c_instrument
10/21/15, 1:44 PM

ZPLSC data may contain acoustic interference from the ADCP instrument subject to real-time clock drift of either instrument.

Id: 2007 By: swhite

CP03ISSM-MFD37-07-ZPLSCC000
10/21/15, 1:44 PM 5/14/16, 9:45 AM

Deployment 3: Multi-frequency transducer failure; Post-recovery evaluation of the instrument revealed failures or defects affecting the 125 kHz, 200 kHz and 455 kHz transducers. The precise time of failure cannot be determined and it is unknown to what extent data are or may be affected.

Id: 2020 By: swhite
Flag: suspect Exclude: No

CP03ISSM-MFD37-07-ZPLSCC000
5/28/16, 10:43 AM 10/15/16, 8:14 AM

Deployment 4: 125 kHz transducer failure; Post-recovery evaluation of the instrument revealed failures or defects affecting the 125 kHz transducer. The precise time of failure cannot be determined and it is unknown to what extent data are or may be affected.

Id: 2021 By: swhite
Flag: suspect Exclude: No

CP03ISSM-MFD37
2/1/17, 9:00 PM 3/29/17, 10:00 PM

All seafloor Multi-Function Node (MFD35 and MFD37) instruments have ceased data telemetry because the seafloor multi-function node had a high voltage ground fault and has been shut down to protect the instruments.

Id: 570 By: leila

CP03ISSM-MFD37
12/20/17, 4:00 AM 3/28/18, 12:56 PM

Deployment 7: Multi-Function Node failed.

Id: 335 By: leila
Flag: not_operational Exclude: No

CP03ISSM-MFD37-07-ZPLSCC000
3/25/18, 11:27 AM 10/26/18, 3:52 PM

Deployment 8: 125 kHz transducer failure; Post-recovery evaluation of the instrument revealed failures or defects affecting the 125 kHz transducer. The precise time of failure cannot be determined and it is unknown to what extent data are or may be affected.

Id: 2025 By: swhite
Flag: suspect Exclude: No

CP03ISSM-MFD37-07-ZPLSCC000
10/28/18, 11:43 PM 4/6/19, 2:51 PM

Deployment 9: 125 kHz transducer failure; Post-recovery evaluation of the instrument revealed failures or defects affecting the 125 kHz transducer. The precise time of failure cannot be determined and it is unknown to what extent data are or may be affected.

Id: 2028 By: swhite
Flag: suspect Exclude: No

CP03ISSM-MFD37-07-ZPLSCC000
4/7/19, 10:42 AM 10/1/19, 7:58 AM

Deployment 10: 125 kHz transducer failure; Post-recovery evaluation of the instrument revealed failures or defects affecting the 125 kHz transducer. The precise time of failure cannot be determined and it is unknown to what extent data are or may be affected.

Id: 2030 By: swhite
Flag: suspect Exclude: No

CP03ISSM-MFD37
12/26/19, 7:00 PM

Deployment 11: *UPDATED 2020-06-22: As of 2020-05-05 the ISSM MFN is back online in a reduced power usage mode. All MFN instruments will be powered on and sample 4 times a day (as opposed to 6 times a day). Data from this instruments will be telemetered home with the exception of the OPTAA which will only be availabe upon recovery. *UPDATED 2020-04-27: The MFN has been offline since 2019-01-03 and no telemetered data are expected from MFN instruments for the remainder of the deployment.*MFN offline due to power shortage after WT2 failure from 12/27/2019 to 12/30/2019 and again from 1/3/2019 to 1/6/2019. Instruments with external batteries will continue to log data internally.

Id: 1878 By: cdobson

CP03ISSM
1/7/20, 7:00 PM

Deployment 11: UPDATED 2020-04-27: As of 2020-03-12 telemetered data for most instruments are being sent home via Iridium. A gap in telemetered data is expected between 2020-01-08 and 2020-03-12 and will not be backfilled but these data will be available upon recovery.* As of 2020-01-08 both fleet broadband telemetry pathways have failed and data can no longer be telemetered to shore from this mooring. Instruments will continue to log data which will be available upon recovery.

Id: 1889 By: cdobson
Flag: not_operational Exclude: No

Review Notes

No notes yet.

New Note