Spectral Irradiance

Reference Designator
CP03ISSM-RID26-08-SPKIRB000
Review Status
Review Complete
Note
Depth
7m
Class
SPKIR (Spectral Irradiance)
Make / Model
Satlantic / OCR507 ICSW

Dataset Reviews Last processed: 7/9/19, 8:11 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 telemetered 17 15 0 1 0 0 196,332 1 7 / 1 1 2 Complete
2 recovered_host 166 90 0 22 3 55 10,362,641 7 / 3 4 2 Complete
3 recovered_host 207 207 0 0 0 0 3,288,523 1 7 / 4 2 Complete
4 recovered_host 141 141 0 0 0 0 2,037,719 1 7 / 4 2 Complete
5 recovered_host 248 248 0 0 0 0 3,408,144 1 7 / 4 2 Complete
6 recovered_host 140 122 0 0 5 18 1,942,862 1 7 / 4 2 Complete
7 recovered_host 148 142 0 0 1 6 2,280,219 1 7 / 4 2 Complete
Data Ranges Review Images

Test Notes

  1. no other streams for comparison
  2. missing: ['pressure']
  3. pass unique test - ascending not tested
  4. no comparison: timestamps do not match

Data Coverage

Deployment: 1234567
88%
54%100%100%100%87%96%

Lat/Lon Differences (km)

Deployment: 1234567
1 0.00
2 0.640.00
3 0.880.630.00
4 0.620.030.650.00
5 0.040.650.910.630.00
6 0.600.900.640.890.640.00
7 0.690.060.680.070.690.960.00

Example Composite Data Plot

System Annotations

Metadata Start Date End Date Comment
CP03ISSM-RID26-08-SPKIRB000
12/14/14, 1:52 PM 12/29/14, 7:00 PM

No recovered host data available.

Id: 454 By: leila

CP03ISSM-RID26
Method: telemetered
6/17/18, 4:04 PM 10/26/18, 3:52 PM

Data telemetry disabled due to chronic issues with power generation and management. The instruments could be revived to send data via telemetry when sufficient power is available. Note: Instruments may continue to collect data on internal battery power.

Id: 1510 By: leila

CP03ISSM-RID26-08-SPKIRB000
9/26/19, 9:50 AM

The use of UV lamps has been implemented for bio-fouling mitigation on this instrument as of the start date/time of this annotation.

Id: 1774 By: swhite

CP03ISSM
1/7/20, 7:00 PM

Deployment 11: UPDATED 2020-04-27: As of 2020-03-12 telemetered data for most instruments are being sent home via Iridium. A gap in telemetered data is expected between 2020-01-08 and 2020-03-12 and will not be backfilled but these data will be available upon recovery.* As of 2020-01-08 both fleet broadband telemetry pathways have failed and data can no longer be telemetered to shore from this mooring. Instruments will continue to log data which will be available upon recovery.

Id: 1889 By: cdobson
Flag: not_operational Exclude: No

Review Notes

Metadata Start Date End Date Comment
CP03ISSM-RID26-08-SPKIRB000

There is no pressure variable in the .nc files. This should be a coordinate.

By Lori Garzio, on 7/12/19

CP03ISSM-RID26-08-SPKIRB000

A substantial percentage (between 11-37%) of data are outside of global ranges for every deployment. Zooming in on the x-axis in the plots shows that all of the values that are outside of global ranges are negative.

By Lori Garzio, on 7/12/19

CP03ISSM-RID26-08-SPKIRB000

Data gaps for deployments 2, 6 and 7 should be annotated.

By Lori Garzio, on 7/12/19

CP03ISSM-RID26-08-SPKIRB000
Deployment: 5
4/1/17, 12:00 AM 6/15/17, 3:56 PM

The signal at wavelength=443 nm seems anomalously high at the end of this deployment.

By Lori Garzio, on 7/12/19

CP03ISSM-RID26-08-SPKIRB000
Deployment: 6
10/1/17, 12:00 AM 11/1/17, 2:17 PM

Suspect data - signal gradually decreased to almost zero at the end of this deployment.

By Lori Garzio, on 7/12/19