pCO2 Air-Sea

Reference Designator
CP03ISSM-SBD12-04-PCO2AA000
Review Status
Review Complete
Note
Depth
0m
Class
PCO2A (pCO2 Air-Sea)
Make / Model
Pro-Oceanus / pCO2-pro

Dataset Reviews Last processed:

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 recovered_host / Complete
1 recovered_host / Complete
2 recovered_host / Complete
2 recovered_host / Complete
3 recovered_host 207 207 0 0 0 0 49,380 0 / 1 Complete
3 recovered_host 207 207 0 0 0 0 49,368 1 0 / 1 Complete
4 recovered_host 141 141 0 0 0 0 33,575 0 / 1 2 Complete
4 recovered_host 141 141 0 0 0 0 33,573 1 0 / 1 Complete
5 recovered_host 248 248 0 0 0 0 53,010 0 / 1 Complete
5 recovered_host 248 248 0 0 0 0 53,009 1 0 / 1 Complete
6 recovered_host 140 140 0 0 0 0 30,024 0 / 1 Complete
6 recovered_host 140 140 0 0 0 0 30,024 1 0 / 1 Complete
7 recovered_host 148 124 0 0 0 0 26,109 0 / 1 3 2 Complete
7 recovered_host 148 124 0 0 0 0 26,109 1 0 / 1 3 Complete
Data Ranges Review Images

Test Notes

  1. pass unique test - ascending not tested
  2. fail: check ['CO2 Mole Fraction in Atmosphere', 'pCO2 Atmosphere']
  3. fail: data found in another stream (gaps: [1] days: [19])

Data Coverage

Deployment: 1234567
100%100%100%100%84%
100%100%100%100%84%

Lat/Lon Differences (km)

Deployment: 1234567
3 0.00
4 0.650.00
5 0.910.630.00
6 0.640.890.640.00
7 0.680.070.690.960.00

Example Composite Data Plot

System Annotations

Metadata Start Date End Date Comment
CP03ISSM-SBD12-04-PCO2AA000
Method: recovered_host
12/14/14, 1:52 PM 12/29/14, 7:00 PM

No recovered data available because the Instrument was lost.

Id: 439 By: leila
Flag: not_available Exclude: No

CP03ISSM-SBD12-04-PCO2AA000
6/17/18, 8:00 PM 9/9/18, 8:00 PM

Deployment 8: Instrument was disabled 2018-06-18 through 2018-09-10 due to power generation issues on the buoy. No telemetered or recovered data expected for this time period.

Id: 1824 By: cdobson
Flag: not_operational Exclude: No

CP03ISSM-SBD12
Method: telemetered
6/18/18, 1:56 AM 10/26/18, 3:52 PM

Data telemetry disabled due to chronic issues with power generation and management. The instruments could be revived to send data via telemetry when sufficient power is available. Note: Instruments may continue to collect data on internal battery power.

Id: 1508 By: leila

CP03ISSM-SBD12-04-PCO2AA000
4/7/19, 10:42 AM 9/26/19, 9:50 AM

Due to a firmware bug, the sensor zeroing function for correcting the baseline drift over time is not currently saving the new reference value to compare the raw CO2 value outputs during the measurements. The sensors are zeroing at the end of the sample cycle when scheduled, but the firmware is not storing the new value. This results in the original zero reference value being used for all measurements and an observable and consistent negative drift in both the air and water CO2 values. The vendor has indicated that the data may be correctable post-recovery, however at this time no such correction algorithm has been provided

Id: 1715 By: swhite
Flag: fail Exclude: Yes

CP03ISSM-SBD12-04-PCO2AA000
11/3/19, 7:00 PM 11/14/19, 7:00 AM

Deployment 11: *UPDATED 2019-11-14: Instrument is powered back on and functional as of 2019-11-14. A data gap is expected 2019-11-04 through 2019-11-14.* PCO2A2 was disabled 2019-11-04 after issues were observed starting 2019-11-04. No data expected after this point.

Id: 1813 By: cdobson

CP03ISSM
1/7/20, 7:00 PM

Deployment 11: UPDATED 2020-04-27: As of 2020-03-12 telemetered data for most instruments are being sent home via Iridium. A gap in telemetered data is expected between 2020-01-08 and 2020-03-12 and will not be backfilled but these data will be available upon recovery.* As of 2020-01-08 both fleet broadband telemetry pathways have failed and data can no longer be telemetered to shore from this mooring. Instruments will continue to log data which will be available upon recovery.

Id: 1889 By: cdobson
Flag: not_operational Exclude: No

CP03ISSM-SBD12-04-PCO2AA000
3/31/20, 5:59 PM

Deployment 11: The PCO2A instrument failed (possibly a membrane or lamp failure). The instrument was powered down for the rest of the deployment on 4/9/2020.

Id: 2048 By: swhite

Review Notes

Metadata Start Date End Date Comment
CP03ISSM-SBD12-04-PCO2AA000

pco2_co2flux is an array of fill values for every deployment. See CP01CNSM-SBD12-04-PCO2AA000 for additional notes on potentially the same issue.

By Lori Garzio, on 7/31/19

CP03ISSM-SBD12-04-PCO2AA000
Deployment: 1

There are no data available for download for deployment 1. According to annotation ID 439, recovered data aren't available because the instrument was lost. However, according to the ingest csv, raw telemetered data files are available. These files should be ingested or the annotation should be updated to indicate that no data are available for this deployment.

By Lori Garzio, on 7/31/19

CP03ISSM-SBD12-04-PCO2AA000
Deployment: 2

There are no data available for download for this deployment. According to the ingest csv, raw recovered_host data files are available. These files should be ingested or the dataset should be annotated.

By Lori Garzio, on 7/31/19

CP03ISSM-SBD12-04-PCO2AA000
Deployment: 3
2/10/16, 12:00 AM 2/21/16, 12:00 AM

Suspect data - measured_air_co2 values are too high

By Lori Garzio, on 7/31/19

CP03ISSM-SBD12-04-PCO2AA000
Deployment: 4
9/9/16, 12:00 AM 10/15/16, 12:14 PM

Instrument appears to fail - this should be annotated.

By Lori Garzio, on 7/31/19

CP03ISSM-SBD12-04-PCO2AA000
Deployment: 5

pCO2 Atmosphere (uatm) values at the beginning of this deployment are substantially higher than the values at the end of deployment 4.

By Lori Garzio, on 7/31/19