Dep. | Preferred Method | Stream | DD | FD | SG | EG | Gaps | GD | TS | Rate (s) | Pressure Comp. | Time Order | Valid Data | Missing Data | Data Comp. | Missing Coords. | Review |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
1 | recovered_inst | pco2w_abc_instrument | / | Complete | |||||||||||||
2 | recovered_host | 168 | 3 | 0 | 164 | 0 | 0 | 34 | 3,599 | 442 / | 1 | Complete | |||||
5 | recovered_host | 240 | 5 | 0 | 234 | 0 | 0 | 47 | 7,199 | 450 / | 1 | Complete | |||||
6 | recovered_host | 150 | 22 | 0 | 127 | 0 | 0 | 422 | 3,599 | 443 / | 1 | Complete | |||||
7 | telemetered | 153 | 105 | 0 | 47 | 0 | 0 | 2,213 | 3,600 | 452 / | 2 | 2 | 1 | Complete |
Deployment: | 1 | 2 | 5 | 6 | 7 |
---|---|---|---|---|---|
69% | |||||
2% | 2% | 15% |
Deployment: | 1 | 2 | 5 | 6 | 7 |
---|---|---|---|---|---|
2 | 0.00 | ||||
5 | 0.82 | 0.00 | |||
6 | 0.20 | 0.91 | 0.00 | ||
7 | 0.75 | 0.08 | 0.84 | 0.00 |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
CP04OSSM-MFD35-05-PCO2WB000 Method: telemetered Stream: pco2w_abc_dcl_instrument |
12/12/14, 4:47 PM | 1/9/19, 7:00 PM |
* UPDATED, 2019-01-25: After careful review, the original data product algorithm (pco2_pco2wat) for this sensor was found to contain an error that resulted in calculation of incorrect derived L1 pCO2 values. In consultation with the vendor, the algorithm has been corrected, verified, and uploaded to the OOI system. This update, effective 2019-01-10, applies to all deployments for this sensor. All derived L1 Id: 293 By: friedrich.knuth |
CP04OSSM-MFD35-05-PCO2WB000 Method: recovered_host Stream: pco2w_abc_dcl_instrument_recovered |
12/12/14, 4:47 PM | 1/9/19, 7:00 PM |
* UPDATED, 2019-01-25: After careful review, the original data product algorithm (pco2_pco2wat) for this sensor was found to contain an error that resulted in calculation of incorrect derived L1 pCO2 values. In consultation with the vendor, the algorithm has been corrected, verified, and uploaded to the OOI system. This update, effective 2019-01-10, applies to all deployments for this sensor. All derived L1 Id: 302 By: friedrich.knuth |
CP04OSSM Method: telemetered |
10/22/15, 3:38 PM | 5/13/16, 1:18 PM |
Fuel cell impacted available instrument power, so telemetered data are not available or reliable. Recovered data are available for instruments that could record internally. Id: 1362 By: leila |
CP04OSSM-MFD35 |
10/16/16, 8:00 PM | 4/2/17, 9:00 PM |
No telemetered data are expected because of ground faults on the seafloor Multi-Function Node. Recovered data are available from instruments recording internally. Id: 355 By: leila |
CP04OSSM-MFD35 Method: telemetered |
2/9/18, 1:00 PM | 3/29/18, 8:42 AM |
Deployment 7: Communication with the MFN instruments was disabled due to intermittent communications over the DSL line.
Id: 937
By: leila
|
CP04OSSM-MFD35 Method: telemetered |
5/9/18, 11:00 PM | 10/30/18, 10:30 AM |
Instruments are presently disabled. Note: Instruments may continue to sample using internal batteries and storage cards. Id: 1517 By: leila |
CP04OSSM Method: telemetered |
1/30/19, 10:00 AM | 4/8/19, 7:39 AM |
CP04OSSM D00009 buoy broke free and went adrift on 1/30/2019, 1500UTC. All instruments on the MFN ceased telemetry on 1/30/2019 and instruments on the buoy/NSIF continued to sample through 2/1/2019. No more telemetered data expected for this deployment. Id: 1557 By: cdobson |
CP04OSSM-MFD35 |
5/11/19, 8:00 PM |
Deployment 10: Telemetered data from all MFN instruments could be delayed up to 3 days due to a limitation in communications with the MFN. Id: 1628 By: cdobson |
|
CP04OSSM-MFD35 |
11/14/19, 7:00 PM |
Deployment 11: MFN offline due to high voltage short on 2019-11-15. No telemetered data expected from instruments on the MFN; instruments with external batteries will continue to log internally.
Id: 1821
By: cdobson
|
Metadata | Start Date | End Date | Comment |
---|---|---|---|
CP04OSSM-MFD35-05-PCO2WB000 |
There are no recovered_inst data available for download for any deployment. According to the ingest csvs, the recovered_inst raw data files are available for deployments 2, and 5-7. These files should be ingested or the datasets should be annotated to explain why the recovered_inst data aren't available. By Lori Garzio, on 7/25/19 |
||
CP04OSSM-MFD35-05-PCO2WB000 |
There is no pressure variable in the .nc files. This should be a coordinate. By Lori Garzio, on 7/25/19 |
||
CP04OSSM-MFD35-05-PCO2WB000 |
Data generally look suspect for many deployments because of the high percentage of values outside of global ranges. The dataset should be annotated if there is a known issue with this instrument to inform users if all or some data should be considered suspect. By Lori Garzio, on 7/25/19 |
||
CP04OSSM-MFD35-05-PCO2WB000 Deployment: 1 |
There are no data available for download for this deployment. According to the ingest csv, the recovered_host raw data files are available (recovered_inst files are not expected). The recovered_host files should be ingested or the dataset should be annotated to explain why the data aren't available. By Lori Garzio, on 7/25/19 |
||
CP04OSSM-MFD35-05-PCO2WB000 Deployment: 2 |
There are only 2 days of data and all of the values for pco2_seawater are nans. By Lori Garzio, on 7/25/19 |
||
CP04OSSM-MFD35-05-PCO2WB000 Deployment: 5 |
|||
CP04OSSM-MFD35-05-PCO2WB000 Deployment: 6 |
|||
CP04OSSM-MFD35-05-PCO2WB000 Deployment: 7 |