No reviews found.
Metadata | Start Date | End Date | Comment |
---|---|---|---|
CP05MOAS-GL375 |
1/11/15, 4:47 PM | 2/12/15, 7:00 PM |
Glider lost. No recovered data are available. Id: 673 By: leila |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
CP05MOAS |
Bad altimeter pings are apparent in many deployments of the coastal gliders. A few example plots are attached - m_water_depth plotted as semi-transparent bathymetry. By Lori Garzio, on 5/10/19 |
||
CP05MOAS |
Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue. By Lori Garzio, on 5/15/19 |