3-Wavelength Fluorometer

Reference Designator
CP05MOAS-GL376-02-FLORTM000
Review Status
Review Complete
Note
Depth
0 to 1,000m
Class
FLORT (3-Wavelength Fluorometer)
Make / Model
WET Labs / ECO Puck FLBBCD-SLK

Dataset Reviews Last processed: 4/15/19, 7:12 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 recovered_host flort_m_sample 69 68 0 0 0 0 2,326,051 1 1000 / 986 1 Complete
2 telemetered flort_m_sample 9 7 0 0 0 0 8,494 60 1000 / 177 2 2 Complete
3 recovered_host flort_m_sample 58 51 0 6 0 0 902,599 1 1000 / 988 3 Complete
4 recovered_host flort_m_sample 13 7 0 4 1 1 181,835 1 1000 / 984 Complete
5 recovered_host flort_m_sample 75 75 0 0 0 0 2,693,572 1 1000 / 1000 4 Complete
7 recovered_host flort_m_sample 72 72 0 0 0 0 2,622,774 1 1000 / 996 5 Complete
8 recovered_host flort_m_sample 92 92 0 0 0 0 3,288,297 1 1000 / 968 6 Complete
Data Ranges Review Images

Test Notes

  1. fail: data found in another stream (gaps: [5] days: [5])
  2. no other streams for comparison
  3. fail: data found in another stream (gaps: [1] days: [1])
  4. fail: data found in another stream (gaps: [4] days: [4])
  5. fail: data found in another stream (gaps: [3] days: [3])
  6. fail: data found in another stream (gaps: [7] days: [7])

Data Coverage

Deployment: 1234578
flort_m_sample 99%78%88%54%100%100%100%

Deployment Map

System Annotations

Metadata Start Date End Date Comment
CP05MOAS-GL376
5/6/15, 7:57 AM 5/13/15, 9:55 PM

Deployment 2: Glider experienced an air bladder leak shortly after deployment and needed to be recovered; no data are expected for this deployment.

Id: 2039 By: cdobson

CP05MOAS-GL376
1/21/16, 10:15 AM 2/2/16, 9:06 AM

Deployment 4: Glider lost its rudder in a storm and could no longer steer properly so it was recovered early.

Id: 2040 By: cdobson

CP05MOAS-GL376
4/19/17, 4:43 PM 4/19/17, 11:30 PM

Deployment 6: Glider was deployed for a test deployment and recovered shortly after. No data are expected for this deployment.

Id: 2041 By: cdobson

CP05MOAS-GL376
Method: telemetered
9/25/18, 6:00 PM 10/1/18, 9:00 AM

Deployment 9: Glider was recovered early on 2018-10-01 1300 UTC. The glider sustained damage to the rudder during a routine surface interval on 2018-09-26 and subsequently lost steering control. The root cause is not yet known.

Id: 1532 By: cdobson

CP05MOAS-GL376
6/18/19, 7:36 PM 6/26/19, 2:45 PM

Deployment 10: Glider was experiencing telemetry issues and needed to be recovered early.

Id: 2042 By: cdobson

Review Notes

Metadata Start Date End Date Comment
CP05MOAS

Bad altimeter pings are apparent in many deployments of the coastal gliders. A few example plots are attached - m_water_depth plotted as semi-transparent bathymetry.

By Lori Garzio, on 5/10/19

CP05MOAS

Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue.

By Lori Garzio, on 5/15/19

CP05MOAS-GL376-02-FLORTM000

Comparisons with the shipboard CTD casts were done for D0001, D0003, D0005, D0007, and D0008. Chl-a recorded by the glider was higher compared to the CTD cast (particularly at the chl-a maximum). Further analysis using the shipboard bottle data should be conducted. This is outside the scope of this project.

By Lori Garzio, on 3/13/19

CP05MOAS-GL376
Deployment: 1
11/15/14, 12:00 AM 12/13/14, 12:00 AM

The data at the end of the deployment (when the glider was drifting/stuck at the surface?) should be annotated as suspect.

By Lori Garzio, on 5/7/19

CP05MOAS-GL376-02-FLORTM000
Deployment: 3

85% of the chl-a data are negative. This should be annotated until the issue is resolved.

By Lori Garzio, on 5/8/19

CP05MOAS-GL376
Deployment: 3
11/15/15, 12:00 AM 11/30/15, 12:00 AM

The sampling pattern for all instruments changed on 11/15/2015. This should be annotated.

By Lori Garzio, on 5/7/19

CP05MOAS-GL376
Deployment: 4

Glider doesn't look like it sampled according to plan. This dataset should be annotated.

By Lori Garzio, on 5/7/19

CP05MOAS-GL376-02-FLORTM000
Deployment: 4

99% of the chl-a data are negative. This should be annotated until the issue is resolved.

By Lori Garzio, on 5/8/19

CP05MOAS-GL376-02-FLORTM000
Deployment: 4

Suspect data >400 dbar in the beginning half of the deployment.

By Lori Garzio, on 5/8/19

CP05MOAS-GL376-02-FLORTM000
Deployment: 5

47% of the chl-a data are negative. This should be annotated until the issue is resolved.

By Lori Garzio, on 5/8/19

CP05MOAS-GL376-02-FLORTM000
Deployment: 5

Suspect data >800 dbar at the end of the deployment.

By Lori Garzio, on 5/8/19

CP05MOAS-GL376-02-FLORTM000
Deployment: 8

Lots of noise in the data through the water column, particularly between 11/8/2017 - 11/20/2017

By Lori Garzio, on 5/8/19