Metadata Start Date End Date Comment
CP05MOAS-GL379
1/16/17, 7:46 AM 1/16/17, 6:59 PM

Glider leaked and was recovered within 12 hours. No recovered data are expected.

Id: 357 By: leila
Flag: not_available Exclude: No

CP05MOAS-GL379
1/16/17, 7:46 AM 1/16/17, 6:59 PM

Glider recovered same day for leak.

Id: 643 By: leila
Flag: not_available Exclude: No

CP05MOAS-GL379
11/18/19, 3:59 PM 11/18/19, 11:35 PM

Deployment 8: Glider experienced early battery failure and science was disabled on 2019-11-18. No telemetered or recovered science data expected after this date.

Id: 1822 By: cdobson
Flag: not_operational Exclude: No

Review Notes

Metadata Start Date End Date Comment
CP05MOAS

Bad altimeter pings are apparent in many deployments of the coastal gliders. A few example plots are attached - m_water_depth plotted as semi-transparent bathymetry.

By Lori Garzio, on 5/10/19

CP05MOAS

Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue.

By Lori Garzio, on 5/15/19

CP05MOAS-GL379
Deployment: 4

Annotation ID 643 is a duplicate of 357

By Lori Garzio, on 5/7/19