Mobile Asset Controller

Reference Designator
CP05MOAS-GL387-00-ENG000000
Review Status
Engineering
Note
Depth
0 to 1,000m
Class
ENG00 (Mobile Asset Controller)
Make / Model
Teledyne Webb / G2 Slocum Gliders

Dataset Reviews

No reviews found.

Deployment Map

System Annotations

Metadata Start Date End Date Comment
CP05MOAS-GL387
7/13/16, 9:00 PM 7/28/16, 8:00 AM

The glider altimeter may have failed.

Id: 644 By: leila
Flag: not_available Exclude: No

CP05MOAS-GL387
5/26/19, 8:00 PM

Deployment 7: Glider's altimeter failed on 5/27/2019 so it will be limited to sampling the bottom line of the FZ box and to 500m until the end of the deployment.

Id: 1620 By: cdobson

Review Notes

Metadata Start Date End Date Comment
CP05MOAS

Bad altimeter pings are apparent in many deployments of the coastal gliders. A few example plots are attached - m_water_depth plotted as semi-transparent bathymetry.

By Lori Garzio, on 5/10/19

CP05MOAS

Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue.

By Lori Garzio, on 5/15/19

CP05MOAS-GL387
Deployment: 1

Glider doesn't look like it sampled according to plan. This should be annotated.

By Lori Garzio, on 9/5/19

New Note