3-Wavelength Fluorometer

Reference Designator
CP05MOAS-PG564-03-FLORTM000
Review Status
Review Complete
Note
Depth
0 to 200m
Class
FLORT (3-Wavelength Fluorometer)
Make / Model
WET Labs / ECO Puck FLBBCD-SLK

Dataset Reviews Last processed: 5/29/19, 6:48 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
2 recovered_host flort_m_sample 52 48 0 4 0 0 458,738 1 200 / 197 Complete
Data Ranges Review Images

Test Notes

All good!

Data Coverage

Deployment: 2
flort_m_sample 92%

Deployment Map

System Annotations

No annotations yet.

Review Notes

Metadata Start Date End Date Comment
CP05MOAS

Bad altimeter pings are apparent in many deployments of the coastal gliders. A few example plots are attached - m_water_depth plotted as semi-transparent bathymetry.

By Lori Garzio, on 5/10/19

CP05MOAS

Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue.

By Lori Garzio, on 5/15/19

CP05MOAS-PG564

There are no global ranges in the system for any science parameter from all instruments on this glider.

By Lori Garzio, on 5/30/19

CP05MOAS-PG564-03-FLORTM000
Deployment: 2

Suspect backscatter profile at the end of the deployment.

By Lori Garzio, on 5/30/19