Hydrogen Sensor

Reference Designator
GA01SUMO-SBD11-03-HYDGN0000
Review Status
Engineering
Note
Depth
0m
Class
HYDGN (Hydrogen Sensor)
Make / Model
RKI Instruments / M2A transmitters

Dataset Reviews

No reviews found.

System Annotations

Metadata Start Date End Date Comment
GA01SUMO
3/15/15, 5:25 PM 11/26/15, 5:08 AM

Deployment 1: Buoy log files showed that seawater was first detected inside the buoy well on 3/26/2015. Both leak detectors showed increasing presence of water over the period of deployment. Upon recovery of the mooring, no external damage was visible and the buoy did not appear to be floating lower than normal.

Id: 653 By: lgarzio

GA01SUMO-SBD11
3/15/15, 5:25 PM 11/26/15, 5:08 AM

Deployment 1: Upon recovery, the bottom of the buoy was heavily biofouled with gooseneck barnacles. The barnacle coverage likely caused imporper sea water flushing at the location of the intakes and sensor faces for the subsurface instruments.

Id: 654 By: lgarzio

GA01SUMO-SBD11
11/14/15, 4:06 PM 11/8/16, 5:10 AM

Deployment 2: Upon recovery, the bottom of the buoy was heavily biofouled with barnacles.

Id: 663 By: lgarzio

GA01SUMO-SBD11
10/26/16, 9:47 PM 1/14/18, 5:24 AM

Deployment 3: Upon recovery, the bottom of the buoy was heavily biofouled.

Id: 385 By: lgarzio

Review Notes

Metadata Start Date End Date Comment
GA01SUMO-SBD11-03-HYDGN0000
Deployment: 1
3/15/15, 12:00 AM 11/26/15, 12:00 AM

Recovered data were not archived and are not available.

By Lori Garzio, on 3/16/17
Redmine Issue #10637

New Note