GA01SUMO
3/15/15, 5:25 PM 11/26/15, 5:08 AM

Deployment 1: Buoy log files showed that seawater was first detected inside the buoy well on 3/26/2015. Both leak detectors showed increasing presence of water over the period of deployment. Upon recovery of the mooring, no external damage was visible and the buoy did not appear to be floating lower than normal.

Id: 653 By: lgarzio

GA01SUMO-SBD12
3/15/15, 5:25 PM 11/26/15, 5:08 AM

Deployment 1: Upon recovery, the bottom of the buoy was heavily biofouled with gooseneck barnacles. The barnacle coverage likely caused imporper sea water flushing at the location of the intakes and sensor faces for the subsurface instruments.

Id: 655 By: lgarzio

GA01SUMO-SBD12
11/14/15, 4:06 PM 11/8/16, 5:10 AM

Deployment 2: Upon recovery, the bottom of the buoy was heavily biofouled with barnacles.

Id: 664 By: lgarzio

GA01SUMO-SBD12
10/26/16, 9:47 PM 1/14/18, 5:24 AM

Deployment 3: Upon recovery, the bottom of the buoy was heavily biofouled.

Id: 386 By: lgarzio

Review Notes

Metadata Start Date End Date Comment
GA01SUMO-SBD12-03-HYDGN0000
Deployment: 1
3/15/15, 12:00 AM 11/26/15, 12:00 AM

Recovered data were not archived and are not available.

By Lori Garzio, on 3/16/17
Redmine Issue #10637

New Note