Dissolved Oxygen

Reference Designator
GA05MOAS-GL364-02-DOSTAM000
Review Status
Review Complete
Note
Depth
0 to 1,000m
Class
DOSTA (Dissolved Oxygen Stable Response)
Make / Model
Aanderaa / Optode 4831

Dataset Reviews Last processed: 4/10/19, 6:38 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
2 telemetered 434 327 0 107 0 0 95,065 60 1000 / 1001 1 1 Complete
Data Ranges Review Images

Test Notes

  1. no other streams for comparison

Data Coverage

Deployment: 2
75%

Deployment Map

System Annotations

Metadata Start Date End Date Comment
GA05MOAS-GL364
10/31/16, 8:58 PM 9/22/17, 9:00 PM

Glider was blown out of the array due to strong currents, went adrift and stopped collecting science data on 9/23/2017. Glider was not recovered - only telemetered data are available.

Id: 436 By: lgarzio

Review Notes

Metadata Start Date End Date Comment
GA05MOAS

Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue.

By Lori Garzio, on 5/31/19

New Note