Dep. | Preferred Method | Stream | DD | FD | SG | EG | Gaps | GD | TS | Rate (s) | Pressure Comp. | Time Order | Valid Data | Missing Data | Data Comp. | Missing Coords. | Review |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
1 | telemetered | flort_m_sample | 313 | 210 | 21 | 82 | 0 | 0 | 75,985 | 15 | 200 / 211 | 1 | 1 | Complete |
Deployment: | 1 |
---|---|
flort_m_sample | 67% |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
GA05MOAS-PG578 |
10/31/16, 9:58 PM | 6/18/17, 11:55 PM |
Evidence of biofouling: caused degraded steering and slow ascent rates. Glider went adrift and stopped collecting science data on 6/19/2017. Glider was not recovered - only telemetered data are available. Id: 171 By: lgarzio |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
GA05MOAS |
Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue. By Lori Garzio, on 5/31/19 |
||
GA05MOAS-PG578 Deployment: 1 |
|||
GA05MOAS-PG578-03-FLORTM000 Deployment: 1 |