Dep. | Preferred Method | Stream | DD | FD | SG | EG | Gaps | GD | TS | Rate (s) | Pressure Comp. | Time Order | Valid Data | Missing Data | Data Comp. | Missing Coords. | Review |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
1 | recovered_host | 347 | 20 | 0 | 327 | 1 | 1 | 7,671 | 12 / 12 | 1 | Complete | ||||||
2 | recovered_inst | 339 | 1 | 26 | 311 | 0 | 0 | 473 | 12 / 12 | 2 | Complete | ||||||
3 | recovered_inst | 401 | 64 | 29 | 122 | 22 | 196 | 2,474 | 12 / 12 | 3 | 2 | Complete | |||||
4 | recovered_inst | 314 | 314 | 0 | 0 | 0 | 0 | 30,007 | 900 | 12 / 4 | 2 | Complete |
Deployment: | 1 | 2 | 3 | 4 |
---|---|---|---|---|
6% | ||||
0% | 16% | 100% |
Deployment: | 1 | 2 | 3 | 4 |
---|---|---|---|---|
1 | 0.00 | |||
2 | 5.69 | 0.00 | ||
3 | 0.37 | 6.04 | 0.00 | |
4 | 1.39 | 5.60 | 1.39 | 0.00 |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
GI01SUMO-RID16-03-CTDBPF000 |
9/10/14, 2:50 PM | 8/22/15, 7:00 AM |
Deployment 1: The Y-cable attached to the pump and the NSIF was missing at recovery. No recovered_instrument data are available. Id: 123 By: lgarzio |
GI01SUMO |
2/14/15, 7:00 PM | 8/21/15, 8:00 PM |
Deployment 1: Status data sent from the buoy included leak detects in the buoy well, drop in battery voltage, and loss of wind turbine input. Upon recovery, the buoy was primarily intact but several instruments were damaged and/or missing. Ice build-up on the tower is speculated to be the cause for much of the damage. No telemetered or recovered_host data expected. Functional instruments could continue to collect data using internal battery power and storage cards. Id: 46 By: lgarzio |
GI01SUMO-RID16-03-CTDBPF000 Method: telemetered Stream: ctdbp_cdef_dcl_instrument |
6/11/15, 12:00 PM | 6/29/15, 11:20 PM |
Telemetry was re-established after suspected icing over of the surface buoy. Data fail basic QC tests.
Id: 124
By: lgarzio
|
GI01SUMO-RID16-03-CTDBPF000 |
8/15/15, 3:22 PM | 7/18/16, 8:00 PM |
Deployment 2: Instrument did not record valid data for the majority of the deployment.
Id: 122
By: lgarzio
|
GI01SUMO |
1/26/16, 7:00 PM | 7/18/16, 8:00 PM |
Deployment 2: A period of violent weather caused power outages on multiple instruments. No telemetered or recovered_host data expected. Functional instruments could continue to collect data using internal battery power and storage cards. Upon recovery, the buoy well was flooded. Id: 47 By: lgarzio |
GI01SUMO-RID16-03-CTDBPF000 |
7/10/16, 1:44 PM | 8/14/17, 5:33 AM |
Deployment 3: Intermittent sampling, possibly due to a misconfiguration with the buoy. Id: 707 By: lgarzio |
GI01SUMO |
12/26/16, 7:00 PM | 4/6/17, 8:00 PM |
Deployment 3: Wind turbines were disabled due to evidence of ice build-up on the surface buoy tower. The platform's battery charge state declined steadily. As of 2017-01-15, all DCLs were shut down and the telemetry schedule was reduced to conserve power. Battery-powered instruments should continue to collect data. UPDATE 2017-04-07: With increasing daylight at this location, sufficient power was generated by solar panels to re-establish limited high-bandwidth telemetry and data transmission. UPDATE: at recovery, wind turbines were missing their hubs and blades, and one solar panel was missing. Id: 121 By: lgarzio |
GI01SUMO |
8/5/17, 2:17 PM | 6/14/18, 4:02 AM |
Deployment 4: at 10:03 UTC on 12 October 2017 the Irminger Sea Surface Mooring stopped all communications. At recovery, the surface buoy was missing. The subsurface portion of the mooring had fallen to the seafloor and was recovered. CTDMOs, CTDBP-F, DOSTA, and ADCP instrumentation were full ocean depth rated and survived, although any recovered data available after 2017-10-12T10:03:00 are suspect. All other instrumentation imploded or flooded. The cause for the loss of the buoy is inconclusive. Id: 102 By: lgarzio |
GI01SUMO-RID16-03-CTDBPF000 Method: telemetered |
2/5/19, 7:00 PM |
Deployment 5: Instrument unable to succesfully power on and sample since 2/7/2019. Any data after this date are suspect. Instrument may be collecting data internally on battery power, but is unlikely. Id: 1588 By: cdobson |
|
GI01SUMO |
6/19/19, 8:00 PM |
Deployment 5: Battery voltage dropped below 24V. All DCLs were turned off. CPM1 was placed on a wake schedule. Data collection and telemetry will be irregular until voltage rebounds. Id: 1625 By: cdobson |
|
GI01SUMO Method: telemetered |
10/21/19, 8:00 PM |
Deployment 6: *UPDATED 2020-04-27: As of 2020-02-21 data for most instruments are being telemetered again (via Iridium). A gap in telemetered data is expected between 2019-10-22 and 2020-02-21 and will not be backfilled but these data will be available upon recovery.*Both fleet broadband units have failed and no more telemetered data is expected at this time. Instruments are believed to be operating as normal and their data will be available upon recovery. Id: 1743 By: cdobson |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
GI01SUMO-RID16-03-CTDBPF000 Deployment: 2 |
There is only a few hours worth of data available for deployment 2. Annotation ID 122 should be updated to indicate what the issue was. By Lori Garzio, on 3/13/19 |
||
GI01SUMO-RID16-03-CTDBPF000 Deployment: 3 |
There are 57 days of data available in the recovered_host data stream that is not available in the recovered_inst data stream towards the end of deployment 3. These data should be available from the recovered_inst (or annotated to indicate why they are not available). By Lori Garzio, on 3/13/19 |
||
GI01SUMO Deployment: 4 |
10/12/17, 12:00 AM | 6/14/18, 8:02 AM |
The surface buoy detached from the mooring and the subsurface portion of the mooring fell to the seafloor (see annotation ID 102). Excluding all mooring data for this time range from final statistics. By Lori Garzio, on 1/15/19 |