Seawater pH (20 m)

Reference Designator
GI01SUMO-RII11-02-PHSENE041
Review Status
Review Complete
Note
Depth
20m
Class
PHSEN (Seawater pH)
Make / Model
Sunburst / SAMI-pH- inductive

Dataset Reviews Last processed: 11/27/18, 4:15 AM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 recovered_host 347 84 0 192 1 71 1,470 3,600 20 / 1 Complete
1 recovered_inst / Complete
2 recovered_inst 339 121 0 218 0 0 2,885 3,600 20 / 1 Complete
3 recovered_host 401 8 0 391 0 0 165 3,600 20 / 1 Complete
3 recovered_inst / Complete
4 telemetered 314 34 0 280 0 0 749 3,600 20 / 2 2 1 Complete
4 recovered_inst / Complete
Data Ranges Review Images

Test Notes

  1. missing: ['pressure']
  2. no other streams for comparison

Data Coverage

Deployment: 1234
11%
24%2%
36%

Lat/Lon Differences (km)

Deployment: 1234
1 0.00
2 5.690.00
3 0.376.040.00
4 1.395.601.390.00

Quicklook Images

System Annotations

Metadata Start Date End Date Comment
GI01SUMO-RII11-02-PHSENE041
9/10/14, 2:50 PM 8/22/15, 7:00 AM

Deployment 1: At recovery, the copper shielding was missing and the instrument was broken into 2 pieces. The connector rods were gone and the connecting cable was severed. Recovered_instrument data not available.

Id: 623 By: lgarzio

GI01SUMO
2/14/15, 7:00 PM 8/21/15, 8:00 PM

Deployment 1: Status data sent from the buoy included leak detects in the buoy well, drop in battery voltage, and loss of wind turbine input. Upon recovery, the buoy was primarily intact but several instruments were damaged and/or missing. Ice build-up on the tower is speculated to be the cause for much of the damage. No telemetered or recovered_host data expected. Functional instruments could continue to collect data using internal battery power and storage cards.

Id: 46 By: lgarzio

GI01SUMO
1/26/16, 7:00 PM 7/18/16, 8:00 PM

Deployment 2: A period of violent weather caused power outages on multiple instruments. No telemetered or recovered_host data expected. Functional instruments could continue to collect data using internal battery power and storage cards. Upon recovery, the buoy well was flooded.

Id: 47 By: lgarzio

GI01SUMO-RII11-02-PHSENE041
7/18/16, 8:06 AM 8/14/17, 5:33 AM

Deployment 3: Instrument stopped sending data via satellite telemetry, and was missing from the mooring upon recovery. No recovered_instrument data are available.

Id: 475 By: lgarzio
Flag: not_operational Exclude: No

GI01SUMO
12/26/16, 7:00 PM 4/6/17, 8:00 PM

Deployment 3: Wind turbines were disabled due to evidence of ice build-up on the surface buoy tower. The platform's battery charge state declined steadily. As of 2017-01-15, all DCLs were shut down and the telemetry schedule was reduced to conserve power. Battery-powered instruments should continue to collect data. UPDATE 2017-04-07: With increasing daylight at this location, sufficient power was generated by solar panels to re-establish limited high-bandwidth telemetry and data transmission. UPDATE: at recovery, wind turbines were missing their hubs and blades, and one solar panel was missing.

Id: 121 By: lgarzio

GI01SUMO
8/5/17, 2:17 PM 6/14/18, 4:02 AM

Deployment 4: at 10:03 UTC on 12 October 2017 the Irminger Sea Surface Mooring stopped all communications. At recovery, the surface buoy was missing. The subsurface portion of the mooring had fallen to the seafloor and was recovered. CTDMOs, CTDBP-F, DOSTA, and ADCP instrumentation were full ocean depth rated and survived, although any recovered data available after 2017-10-12T10:03:00 are suspect. All other instrumentation imploded or flooded. The cause for the loss of the buoy is inconclusive.

Id: 102 By: lgarzio

GI01SUMO-RII11-02-PHSENE041
Method: telemetered
9/17/18, 8:00 PM

Deployment 5: Data for this instrument irregular since 9/18/2018 and missing altogether since 10/10/2018.

Id: 1590 By: cdobson

GI01SUMO
6/19/19, 8:00 PM

Deployment 5: Battery voltage dropped below 24V. All DCLs were turned off. CPM1 was placed on a wake schedule. Data collection and telemetry will be irregular until voltage rebounds.

Id: 1625 By: cdobson

GI01SUMO-RII11-02-PHSENE041
8/5/19, 12:00 PM

Deployment 6: Inductive communications to instrument failed upon deployment, no telemetered data expected.

Id: 1709 By: cdobson

GI01SUMO
Method: telemetered
10/21/19, 8:00 PM

Deployment 6: *UPDATED 2020-04-27: As of 2020-02-21 data for most instruments are being telemetered again (via Iridium). A gap in telemetered data is expected between 2019-10-22 and 2020-02-21 and will not be backfilled but these data will be available upon recovery.*Both fleet broadband units have failed and no more telemetered data is expected at this time. Instruments are believed to be operating as normal and their data will be available upon recovery.

Id: 1743 By: cdobson

Review Notes

Metadata Start Date End Date Comment
GI01SUMO-RII11-02-PHSENE041

pressure coordinate missing in the NetCDF data files

By Leila Belabbassi, on 5/29/19

GI01SUMO-RII11-02-PHSENE041
10/25/14, 12:00 AM 7/10/16, 12:00 AM

suspect data

By Leila Belabbassi, on 5/29/19

GI01SUMO-RII11-02-PHSENE041
8/5/17, 12:00 AM 6/14/18, 12:00 AM

suspect data

By Leila Belabbassi, on 5/29/19

GI01SUMO
Deployment: 4
10/12/17, 12:00 AM 6/14/18, 8:02 AM

The surface buoy detached from the mooring and the subsurface portion of the mooring fell to the seafloor (see annotation ID 102). Excluding all mooring data for this time range from final statistics.

By Lori Garzio, on 1/15/19

New Note