Dep. | Preferred Method | Stream | DD | FD | SG | EG | Gaps | GD | TS | Rate (s) | Pressure Comp. | Time Order | Valid Data | Missing Data | Data Comp. | Missing Coords. | Review |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
1 | recovered_host | / | Complete | ||||||||||||||
1 | recovered_host | 347 | 198 | 0 | 148 | 0 | 0 | 4,690 | 0 / | 1 | Complete | ||||||
2 | recovered_host | 340 | 217 | 0 | 39 | 4 | 85 | 5,083 | 0 / | 2 | Complete | ||||||
2 | recovered_host | 340 | 217 | 0 | 39 | 4 | 85 | 5,083 | 0 / | Complete | |||||||
3 | recovered_host | 401 | 236 | 0 | 1 | 3 | 163 | 5,588 | 0 / | 2 | Complete | ||||||
3 | recovered_host | 401 | 236 | 0 | 1 | 3 | 163 | 5,588 | 0 / | 3 | Complete | ||||||
4 | telemetered | 314 | 69 | 0 | 244 | 0 | 0 | 1,622 | 0 / | 4 | 4 | Complete | |||||
4 | telemetered | 314 | 69 | 0 | 244 | 0 | 0 | 1,622 | 0 / | 4 | 4 | Complete |
Deployment: | 1 | 2 | 3 | 4 |
---|---|---|---|---|
22% | ||||
64% | 59% | |||
22% | ||||
57% | 64% | 59% |
Deployment: | 1 | 2 | 3 | 4 |
---|---|---|---|---|
1 | 0.00 | |||
2 | 5.69 | 0.00 | ||
3 | 0.37 | 6.04 | 0.00 | |
4 | 1.39 | 5.60 | 1.39 | 0.00 |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
GI01SUMO-SBD12-05-WAVSSA000 |
9/10/14, 2:50 PM | 10/12/17, 4:07 AM |
The WAVSS instrument should sample at 4 Hz for 20 minutes every hour according to the OOI baseline sampling strategy. However, during deployments 1 thru 4, the instrument was erroneously configured with a sampling period of only 5 minutes.
Id: 1391
By: lgarzio
|
GI01SUMO |
2/14/15, 7:00 PM | 8/21/15, 8:00 PM |
Deployment 1: Status data sent from the buoy included leak detects in the buoy well, drop in battery voltage, and loss of wind turbine input. Upon recovery, the buoy was primarily intact but several instruments were damaged and/or missing. Ice build-up on the tower is speculated to be the cause for much of the damage. No telemetered or recovered_host data expected. Functional instruments could continue to collect data using internal battery power and storage cards. Id: 46 By: lgarzio |
GI01SUMO |
1/26/16, 7:00 PM | 7/18/16, 8:00 PM |
Deployment 2: A period of violent weather caused power outages on multiple instruments. No telemetered or recovered_host data expected. Functional instruments could continue to collect data using internal battery power and storage cards. Upon recovery, the buoy well was flooded. Id: 47 By: lgarzio |
GI01SUMO |
12/26/16, 7:00 PM | 4/6/17, 8:00 PM |
Deployment 3: Wind turbines were disabled due to evidence of ice build-up on the surface buoy tower. The platform's battery charge state declined steadily. As of 2017-01-15, all DCLs were shut down and the telemetry schedule was reduced to conserve power. Battery-powered instruments should continue to collect data. UPDATE 2017-04-07: With increasing daylight at this location, sufficient power was generated by solar panels to re-establish limited high-bandwidth telemetry and data transmission. UPDATE: at recovery, wind turbines were missing their hubs and blades, and one solar panel was missing. Id: 121 By: lgarzio |
GI01SUMO |
8/5/17, 2:17 PM | 6/14/18, 4:02 AM |
Deployment 4: at 10:03 UTC on 12 October 2017 the Irminger Sea Surface Mooring stopped all communications. At recovery, the surface buoy was missing. The subsurface portion of the mooring had fallen to the seafloor and was recovered. CTDMOs, CTDBP-F, DOSTA, and ADCP instrumentation were full ocean depth rated and survived, although any recovered data available after 2017-10-12T10:03:00 are suspect. All other instrumentation imploded or flooded. The cause for the loss of the buoy is inconclusive. Id: 102 By: lgarzio |
GI01SUMO |
6/19/19, 8:00 PM |
Deployment 5: Battery voltage dropped below 24V. All DCLs were turned off. CPM1 was placed on a wake schedule. Data collection and telemetry will be irregular until voltage rebounds. Id: 1625 By: cdobson |
|
GI01SUMO Method: telemetered |
10/21/19, 8:00 PM |
Deployment 6: *UPDATED 2020-04-27: As of 2020-02-21 data for most instruments are being telemetered again (via Iridium). A gap in telemetered data is expected between 2019-10-22 and 2020-02-21 and will not be backfilled but these data will be available upon recovery.*Both fleet broadband units have failed and no more telemetered data is expected at this time. Instruments are believed to be operating as normal and their data will be available upon recovery. Id: 1743 By: cdobson |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
GI01SUMO-SBD12-05-WAVSSA000 |
There are no data available for download for streams wavss_a_dcl_mean_directional_recovered and wavss_a_dcl_motion_recovered for any deployment. These streams should be available or the datasets should be annotated. By Lori Garzio, on 6/28/19 |
||
GI01SUMO-SBD12-05-WAVSSA000 |
There are no global ranges in the system for wavss_a_dcl_statistics mean_spread, wavss_a_dcl_mean_directional_recovered wavss_a_directional_frequency, and wavss_a_dcl_non_directional_recovered wavss_a_non_directional_frequency By Lori Garzio, on 6/28/19 |
||
GI01SUMO-SBD12-05-WAVSSA000 |
8/15/15, 7:22 PM | 6/14/18, 8:02 AM |
According to annotation ID 1391, the instrument was incorrectly configured for deployments 1-4. However, the number of zero crossings for deployment 1 looks reasonable compared to other OOI WAVSS instruments. Excluding deployments 2-4 from final data ranges. The instrument configuration for deployment 1 should be reviewed and the annotation should be updated if deployment 1 is found to be configured correctly. By Lori Garzio, on 6/28/19 |
GI01SUMO-SBD12-05-WAVSSA000 Deployment: 1 |
The data stream wavss_a_dcl_non_directional_recovered isn't available for download for deployment 1. This stream should be available or the dataset should be annotated. By Lori Garzio, on 6/28/19 |
||
GI01SUMO Deployment: 4 |
10/12/17, 12:00 AM | 6/14/18, 8:02 AM |
The surface buoy detached from the mooring and the subsurface portion of the mooring fell to the seafloor (see annotation ID 102). Excluding all mooring data for this time range from final statistics. By Lori Garzio, on 1/15/19 |