Dissolved Oxygen

Reference Designator
GI03FLMA-RIS01-03-DOSTAD000
Review Status
Review Complete
Note
Depth
30m
Class
DOSTA (Dissolved Oxygen Stable Response)
Make / Model
Aanderaa / Optode 4831

Dataset Reviews Last processed: 7/25/19, 7:19 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 recovered_host 341 341 0 0 0 0 32,602 900 26 / 16 Complete
2 recovered_host 335 176 0 89 1 69 16,503 900 26 / 39 Complete
3 recovered_host 397 50 0 347 0 0 4,665 900 30 / 34 1 Complete
4 recovered_host 312 108 0 135 1 69 10,121 900 30 / 38 Complete
Data Ranges Review Images

Test Notes

  1. fail: check ['DO - Pressure Temp Sal Corrected']

Data Coverage

Deployment: 1234
100%53%13%35%

Lat/Lon Differences (km)

Deployment: 1234
1 0.00
2 2.160.00
3 5.675.670.00
4 2.470.345.890.00

Example Composite Data Plot

System Annotations

Metadata Start Date End Date Comment
GI03FLMA
9/12/14, 4:11 PM 8/18/15, 6:41 AM

Deployment 1: Mooring was deployed 10-15m shallower than planned.

Id: 683 By: lgarzio

GI03FLMA-RIS01-03-DOSTAD000
12/12/14, 7:00 PM 8/18/15, 6:41 AM

Deployment 1: Dissolved oxygen noise increased, possibly from biofouling.

Id: 772 By: lgarzio
Flag: suspect Exclude: No

GI03FLMA
8/18/15, 7:48 PM 7/16/16, 8:00 PM

Deployment 2: Mooring was deployed 5-10m deeper than planned.

Id: 684 By: lgarzio

GI03FLMA-RIS01
1/21/16, 7:00 PM 3/28/16, 8:00 PM

Deployment 2: Secondary controller failed - no recovered_host or telemetered data expected.

Id: 689 By: lgarzio

GI03FLMA-RIS01
4/27/16, 8:00 PM 7/16/16, 8:00 PM

Deployment 2: Secondary controller failed - no leak was detected, but the battery for the controller board was drained at recovery. No recovered_host or telemetered data expected.

Id: 690 By: lgarzio

GI03FLMA-RIS01
8/30/16, 2:00 AM 8/12/17, 7:36 AM

Deployment 3: Secondary controller failed - no recovered_host or telemetered data expected.

Id: 694 By: lgarzio

GI03FLMA
8/8/17, 11:59 AM 6/15/18, 3:59 AM

Deployment 4: Mooring was deployed 5-10m deeper than planned.

Id: 1450 By: lgarzio

GI03FLMA-RIS01
Method: telemetered
6/10/18, 9:59 AM 8/11/19, 4:25 PM

Deployment 5: The secondary controller did not establish communications at deployment; there will be no telemetered data from the instruments connected to this controller. The instruments should be logging internally and data should be available at recovery.

Id: 1427 By: cdobson

Review Notes

Metadata Start Date End Date Comment
GI03FLMA-RIS01

Annotation ID 1427 needs to be updated.

By Lori Garzio, on 3/13/19

GI03FLMA-RIS01-03-DOSTAD000
Deployment: 1
12/1/14, 12:00 AM 8/8/15, 10:41 AM

Suspect data - see annotation ID 772

By Lori Garzio, on 8/1/19

GI03FLMA-RIS01-03-DOSTAD000
Deployment: 4
1/30/18, 12:00 AM 6/15/18, 12:00 AM

Suspect data at the end of the deployment.

By Lori Garzio, on 8/1/19

New Note