Seawater pH

Reference Designator
GI03FLMA-RIS01-04-PHSENF000
Review Status
Review Complete
Note
Depth
30m
Class
PHSEN (Seawater pH)
Make / Model
Sunburst / SAMI-pH

Dataset Reviews Last processed: 7/25/19, 7:15 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 recovered_inst 341 340 0 0 0 0 4,074 7,200 26 / 1 Complete
2 recovered_inst 335 333 0 0 0 0 3,996 7,200 26 / 1 Complete
3 recovered_inst 397 396 0 0 0 0 4,746 7,200 30 / 1 Complete
4 recovered_inst 312 311 0 0 0 0 3,724 7,200 30 / 1 Complete
Data Ranges Review Images

Test Notes

  1. missing: ['pressure']

Data Coverage

Deployment: 1234
100%99%100%100%

Lat/Lon Differences (km)

Deployment: 1234
1 0.00
2 2.160.00
3 5.675.670.00
4 2.470.345.890.00

Example Composite Data Plot

System Annotations

Metadata Start Date End Date Comment
GI03FLMA
9/12/14, 4:11 PM 8/18/15, 6:41 AM

Deployment 1: Mooring was deployed 10-15m shallower than planned.

Id: 683 By: lgarzio

GI03FLMA
8/18/15, 7:48 PM 7/16/16, 8:00 PM

Deployment 2: Mooring was deployed 5-10m deeper than planned.

Id: 684 By: lgarzio

GI03FLMA-RIS01
1/21/16, 7:00 PM 3/28/16, 8:00 PM

Deployment 2: Secondary controller failed - no recovered_host or telemetered data expected.

Id: 689 By: lgarzio

GI03FLMA-RIS01
4/27/16, 8:00 PM 7/16/16, 8:00 PM

Deployment 2: Secondary controller failed - no leak was detected, but the battery for the controller board was drained at recovery. No recovered_host or telemetered data expected.

Id: 690 By: lgarzio

GI03FLMA-RIS01
8/30/16, 2:00 AM 8/12/17, 7:36 AM

Deployment 3: Secondary controller failed - no recovered_host or telemetered data expected.

Id: 694 By: lgarzio

GI03FLMA
8/8/17, 11:59 AM 6/15/18, 3:59 AM

Deployment 4: Mooring was deployed 5-10m deeper than planned.

Id: 1450 By: lgarzio

GI03FLMA-RIS01
Method: telemetered
6/10/18, 9:59 AM 8/11/19, 4:25 PM

Deployment 5: The secondary controller did not establish communications at deployment; there will be no telemetered data from the instruments connected to this controller. The instruments should be logging internally and data should be available at recovery.

Id: 1427 By: cdobson

Review Notes

Metadata Start Date End Date Comment
GI03FLMA-RIS01

Annotation ID 1427 needs to be updated.

By Lori Garzio, on 3/13/19

GI03FLMA-RIS01-04-PHSENF000

There is no pressure variable in the .nc files. This should be a coordinate.

By Lori Garzio, on 7/26/19

GI03FLMA-RIS01-04-PHSENF000
Deployment: 1
4/25/15, 12:00 AM 8/18/15, 10:41 AM

Suspect data in the last several months of the deployment.

By Lori Garzio, on 7/26/19

New Note