2-Wavelength Fluorometer

Reference Designator
GI05MOAS-GL469-01-FLORDM000
Review Status
Review Complete
Note
Depth
0 to 1,000m
Class
FLORD (2-Wavelength Fluorometer)
Make / Model
WET Labs / ECO Puck FLBB-SLC

Dataset Reviews Last processed: 4/12/19, 2:23 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 telemetered 9 8 0 0 0 0 3,994 60 1000 / 984 1 1 Complete
Data Ranges Review Images

Test Notes

  1. no other streams for comparison

Data Coverage

Deployment: 1
89%

Deployment Map

System Annotations

Metadata Start Date End Date Comment
GI05MOAS-GL469
9/10/14, 7:00 PM 9/17/14, 8:00 PM

Deployment 1: Recovered early due to issues with communications.

Id: 567 By: lgarzio

GI05MOAS-GL469
9/10/14, 7:00 PM 9/17/14, 8:00 PM

Glider went adrift and was not recovered. Only telemetered data are available.

Id: 161 By: lgarzio

Review Notes

Metadata Start Date End Date Comment
GI05MOAS-GL469-01-FLORDM000

Comparison with the shipboard CTD cast was done for D0001. Chl-a recorded by the glider was higher compared to the CTD cast. Further analysis using the shipboard bottle data should be conducted. This is outside the scope of this project.

By Lori Garzio, on 3/8/19

GI05MOAS

Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue.

By Lori Garzio, on 5/31/19

GI05MOAS-GL469
Deployment: 1

Annotation IDs 161 and 567 are contradictory and need to be cleaned up. The glider track suggests that 567 is correct. If true, 161 should be deleted.

By Lori Garzio, on 5/7/19