No reviews found.
Metadata | Start Date | End Date | Comment |
---|---|---|---|
GI05MOAS-GL485 |
8/17/15, 7:24 AM | 8/23/15, 8:00 PM |
Deployment 2: Deployed on 8/17/2015, recovered on 8/23/2015 at 13:36 UTC for ballast adjustment. Re-deployed at 13:56 UTC. Id: 565 By: lgarzio |
GI05MOAS-GL485 |
8/17/15, 7:24 AM | 11/19/15, 7:00 PM |
Glider went adrift and was not recovered. Only telemetered data are available. Id: 165 By: lgarzio |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
GI05MOAS |
Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue. By Lori Garzio, on 5/31/19 |