Dep. | Preferred Method | Stream | DD | FD | SG | EG | Gaps | GD | TS | Rate (s) | Pressure Comp. | Time Order | Valid Data | Missing Data | Data Comp. | Missing Coords. | Review |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
2 | telemetered | 81 | 80 | 0 | 0 | 0 | 0 | 27,597 | 60 | 1000 / 985 | 1 | 1 | Complete |
Deployment: | 2 |
---|---|
99% |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
GI05MOAS-GL495 |
8/17/15, 10:08 AM | 11/4/15, 7:00 PM |
Glider went adrift and was not recovered. Only telemetered data are available. Id: 166 By: lgarzio |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
GI05MOAS |
Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue. By Lori Garzio, on 5/31/19 |
||
GI05MOAS-GL495-04-CTDGVM000 |
Comparison with the shipboard CTD cast was done for D0002. The available science parameter values and patterns are comparable between the glider and CTD cast, however further analysis using the shipboard bottle data should be conducted. This is outside the scope of this project. By Lori Garzio, on 3/8/19 |
||
GI05MOAS-GL495-04-CTDGVM000 |