No reviews found.
Metadata | Start Date | End Date | Comment |
---|---|---|---|
GI05MOAS-GL559 |
7/9/16, 6:18 AM | 1/8/17, 4:00 PM |
Deployment 1: Glider was damaged and the batteries failed on 1/8/2017. Glider aborted and was set adrift, recovered off the coast of Greenland in June 2017. Id: 568 By: lgarzio |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
GI05MOAS |
Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue. By Lori Garzio, on 5/31/19 |