2-Wavelength Fluorometer

Reference Designator
GP05MOAS-GL276-01-FLORDM000
Review Status
Review Complete
Note
Depth
0 to 1,000m
Class
FLORD (2-Wavelength Fluorometer)
Make / Model
WET Labs / ECO Puck FLBB-SLC

Dataset Reviews Last processed: 3/3/20, 6:25 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 recovered_host flort_m_sample 353 296 0 0 10 59 5,222,911 1 1000 / 999 1 Complete
Data Ranges Review Images

Test Notes

  1. fail: data found in another stream (gaps: [2] days: [4])

Data Coverage

Deployment: 1
flort_m_sample 84%

Deployment Map

System Annotations

Metadata Start Date End Date Comment
GP05MOAS-GL276
6/15/14, 1:30 AM 6/1/15, 8:00 PM

Deployment 1: Deployed on 6/15/2014, recovered on 6/18/2014 for ballast correction and re-deployed at 07:36 UTC.

Id: 590 By: lgarzio

GP05MOAS-GL276-01-FLORDM000
6/15/14, 1:30 AM 6/1/15, 8:00 PM

Deployment 1: A 3-wavelength fluorometer (instrument series FLORTM) was deployed in place of the customary 2-wavelength fluorometer (instrument series FLORDM) for global gliders.

Id: 1476 By: lgarzio

GP05MOAS-GL276-01-FLORDM000
7/23/18, 2:35 AM 7/31/18, 4:00 PM

Deployment 2: At the beginning of the deployment, the glider didn't send any valid fluorometer data. Issue was resolved and fluorometer data should be available for the remainder of the deployment.

Id: 1475 By: lgarzio

GP05MOAS-GL276-01-FLORDM000
7/23/18, 2:35 AM

Deployment 2: A 3-wavelength fluorometer (instrument series FLORTM) was deployed in place of the customary 2-wavelength fluorometer (instrument series FLORDM) for global gliders.

Id: 1477 By: lgarzio

GP05MOAS-GL276
1/27/19, 8:00 AM 2/12/19, 5:00 PM

Glider 276 was suspected to be showing early signs of battery failure so it was set adrift on 1/27/19 and successfully recovered on 2/12/19. No science data is expected after 1/27/19.

Id: 1561 By: cdobson

Review Notes

Metadata Start Date End Date Comment
GP05MOAS

Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue.

By Lori Garzio, on 5/31/19

GP05MOAS-GL276-01-FLORDM000

Deployments 1 and 2: Data streams for this instrument will not match what we have in the database!! A FLORT was deployed (not a FLORD). The data had to be ingested using the FLORTM driver/parser, meaning the data streams are flort_sample.

By Lori Garzio, on 8/20/18
Redmine Issue #13507

GP05MOAS-GL276
Deployment: 1

I downloaded data for this glider for automated analysis on March 3, 2020. There are a substantial amount of gaps in the CTD, DOSTA, and FLORD data that were not there when I had previously downloaded the data on Apr 12, 2019. There are now 103 days of missing CTD data, 91 days of missing DOSTA data, and 59 days of missing FLORD data that were available on Apr 12, 2019. See Redmine ticket 14654 for details.

By Lori Garzio, on 3/13/20

New Note