2-Wavelength Fluorometer

Reference Designator
GP05MOAS-GL363-01-FLORDM000
Review Status
Review Complete
Note
Depth
0 to 1,000m
Class
FLORD (2-Wavelength Fluorometer)
Make / Model
WET Labs / ECO Puck FLBB-SLC

Dataset Reviews Last processed: 3/3/20, 9:54 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 recovered_host 332 292 0 10 1 30 4,756,045 1 1000 / 993 1 Complete
2 recovered_host 6 5 0 0 0 0 185,775 1 1000 / 999 Complete
Data Ranges Review Images

Test Notes

  1. missing: ['pressure', 'lon', 'lat']

Data Coverage

Deployment: 12
88%83%

Deployment Map

System Annotations

Metadata Start Date End Date Comment
GP05MOAS-GL363
6/29/16, 10:30 PM 7/4/16, 6:45 PM

Deployment 2: Glider recovered shortly after deployment due to pump anomalies.

Id: 591 By: lgarzio

Review Notes

Metadata Start Date End Date Comment
GP05MOAS

Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue.

By Lori Garzio, on 5/31/19

GP05MOAS-GL363
Deployment: 1

latitude and longitude are missing from all deployment 1 .nc files (Redmine #14657).

By Lori Garzio, on 3/16/20