Dep. | Preferred Method | Stream | DD | FD | SG | EG | Gaps | GD | TS | Rate (s) | Pressure Comp. | Time Order | Valid Data | Missing Data | Data Comp. | Missing Coords. | Review |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
1 | recovered_host | 217 | 99 | 0 | 117 | 0 | 0 | 910,636 | 2 | 1000 / 984 | 1 | Complete | |||||
2 | recovered_host | 329 | 246 | 0 | 19 | 1 | 64 | 2,574,969 | 2 | 1000 / 992 | Complete |
Deployment: | 1 | 2 |
---|---|---|
46% | 75% |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
GP05MOAS-GL365 |
6/27/16, 11:35 AM | 5/1/17, 6:53 AM |
Deployment 2: Glider recovered by a ship of opportunity (R/V Sikuliaq) due to limited battery power. Science sensors were turned off on 1/27/2017 to conserve battery, then resumed sampling around 4/2/2017 to 5/1/2017. Id: 594 By: lgarzio |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
GP05MOAS |
Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue. By Lori Garzio, on 5/31/19 |
||
GP05MOAS-GL365 Deployment: 1 |
latitude and longitude are missing from all deployment 1 .nc files (Redmine #14657) By Lori Garzio, on 3/16/20 |
||
GP05MOAS-GL365-02-DOSTAM000 Deployment: 1 |
100% of deployment 1 sci_abs_oxygen values are NaNs (because lat and lon are missing) By Lori Garzio, on 3/17/20 |
||
GP05MOAS-GL365 Deployment: 1 |
|||
GP05MOAS-GL365-02-DOSTAM000 Deployment: 2 |
45% of deployment 2 sci_abs_oxygen values are NaNs By Lori Garzio, on 3/17/20 |