Dataset Reviews Last processed: 3/10/20, 3:48 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 recovered_host 392 279 0 53 8 65 4,911,886 1 1000 / 751 Complete
Data Ranges Review Images

Test Notes

All good!

Data Coverage

Deployment: 1
71%

Deployment Map

System Annotations

Metadata Start Date End Date Comment
GP05MOAS-GL537
6/2/15, 7:21 AM 5/4/16, 6:00 PM

Deployment 1: Glider suffered progressive steering failure and went adrift in May 2016. Digifin was inoperable at the time of recovery, small chips on wings, moderate biofouling.

Id: 589 By: lgarzio

GP05MOAS-GL537
9/29/19, 9:40 PM 9/30/19, 12:01 PM

Deployment 2: Glider was temporarily recovered after its first dive to adjust the external ballast and was re-deployed 2019-09-30 16:01 UTC.

Id: 1764 By: cdobson

GP05MOAS-GL537
4/22/20, 8:00 PM 6/6/20, 11:09 PM

Deployment 2: This glider is being flown to the Oregon coast due to the possibility of limited recovery options at the Papa Array this summer. The glider left the array on 2020-04-23 and the science sensors will continue sampling as it flies toward the coast.

Id: 2071 By: cdobson

Review Notes

Metadata Start Date End Date Comment
GP05MOAS

Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue.

By Lori Garzio, on 5/31/19

New Note