3-Wavelength Fluorometer

Reference Designator
GP05MOAS-PG514-03-FLORTM000
Review Status
Review Blocked
Note
Problem with Raw Rata files
Depth
0 to 200m
Class
FLORT (3-Wavelength Fluorometer)
Make / Model
WET Labs / ECO Puck FLBBCD-SLK

Dataset Reviews

No reviews found.

Deployment Map

System Annotations

Metadata Start Date End Date Comment
GP05MOAS-PG514
6/2/15, 6:00 AM 7/25/15, 2:12 PM

Deployment 2: Recovered data are not available for the beginning of the deployment due to a logging issue with the recovered high-resolution files.

Id: 593 By: lgarzio

GP05MOAS-PG514
6/2/15, 6:00 AM 11/15/15, 7:00 PM

Deployment 2: Glider was found damaged upon recovery, and the NUTNR sensor was missing.

Id: 175 By: lgarzio

Review Notes

Metadata Start Date End Date Comment
GP05MOAS

Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue.

By Lori Garzio, on 5/31/19

New Note