Nitrate

Reference Designator
GP05MOAS-PG514-05-NUTNRM000
Review Status
Review Complete
Note
Depth
0 to 200m
Class
NUTNR (Nitrate)
Make / Model
Satlantic / SUNA V2

Dataset Reviews Last processed: 3/12/20, 2:20 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
2 telemetered 168 167 0 0 0 0 36,964 200 / 394 1 1 Complete
Data Ranges Review Images

Test Notes

  1. no other streams for comparison

Data Coverage

Deployment: 2
99%

Deployment Map

System Annotations

Metadata Start Date End Date Comment
GP05MOAS-PG514
6/2/15, 6:00 AM 7/25/15, 2:12 PM

Deployment 2: Recovered data are not available for the beginning of the deployment due to a logging issue with the recovered high-resolution files.

Id: 593 By: lgarzio

GP05MOAS-PG514
6/2/15, 6:00 AM 11/15/15, 7:00 PM

Deployment 2: Glider was found damaged upon recovery, and the NUTNR sensor was missing.

Id: 175 By: lgarzio

GP05MOAS-PG514-05-NUTNRM000
6/2/15, 6:00 AM 11/15/15, 7:00 PM

Recovered data are not available as the sensor was missing upon recovery.

Id: 176 By: lgarzio

Review Notes

Metadata Start Date End Date Comment
GP05MOAS

Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue.

By Lori Garzio, on 5/31/19

GP05MOAS-PG514-05-NUTNRM000
Deployment: 2

Nitrate data extend to ~1000 dbar at the beginning of the deployment - this glider should only go to 200m. This dataset needs to be annotated if there was a mistake in operations.

By Lori Garzio, on 3/19/20

New Note