3-Wavelength Fluorometer

Reference Designator
GP05MOAS-PG515-03-FLORTM000
Review Status
Review Blocked
Note
Problem with Raw Rata files
Depth
0 to 200m
Class
FLORT (3-Wavelength Fluorometer)
Make / Model
WET Labs / ECO Puck FLBBCD-SLK

Dataset Reviews

No reviews found.

Deployment Map

System Annotations

Metadata Start Date End Date Comment
GP05MOAS-PG515
11/30/15, 7:00 PM 2/16/16, 7:00 PM

Deployment 2: Glider steering failure linked to a leak in the digifin. Frequent motor faults reported beginning in Dec 2015 which lead to disabled dynamic steering control of the glider.

Id: 586 By: lgarzio

Review Notes

Metadata Start Date End Date Comment
GP05MOAS

Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue.

By Lori Garzio, on 5/31/19

New Note