Photosynthetically Available Radiation

Reference Designator
GP05MOAS-PG576-06-PARADM000
Review Status
Will Not Review
Note
Failed
Depth
0 to 200m
Class
PARAD (Photosynthetically Available Radiation)
Make / Model
Biospherical Instruments / QSP-2155

Dataset Reviews

No reviews found.

Deployment Map

System Annotations

No annotations yet.

Review Notes

Metadata Start Date End Date Comment
GP05MOAS

Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue.

By Lori Garzio, on 5/31/19

GP05MOAS-PG576
Deployment: 1
7/5/16, 12:00 AM 7/5/16, 12:00 AM

Deployment 1 - glider deployed on 06/27/2016 and recovered on 07/5/2016 due to a leak in the forward compartment.

By Lori Garzio, on 8/1/16
Redmine Issue #10848

New Note