Dissolved Oxygen

Reference Designator
GS05MOAS-GL484-02-DOSTAM000
Review Status
Review Complete
Note
Depth
0 to 1,000m
Class
DOSTA (Dissolved Oxygen Stable Response)
Make / Model
Aanderaa / Optode 4831

Dataset Reviews Last processed: 4/12/19, 2:25 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 telemetered 2 2 0 0 0 0 382 60 1000 / 1005 1 1 Complete
Data Ranges Review Images

Test Notes

  1. no other streams for comparison

Data Coverage

Deployment: 1
100%

Deployment Map

System Annotations

No annotations yet.

Review Notes

Metadata Start Date End Date Comment
GS05MOAS

Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue.

By Lori Garzio, on 5/31/19

GS05MOAS-GL484
Deployment: 1

Short (2 day) deployment and only telemetered data available for download - dataset should be annotated.

By Lori Garzio, on 5/30/19

New Note