2-Wavelength Fluorometer

Reference Designator
GS05MOAS-GL560-01-FLORDM000
Review Status
Review Complete
Note
Depth
0 to 1,000m
Class
FLORD (2-Wavelength Fluorometer)
Make / Model
WET Labs / ECO Puck FLBB-SLC

Dataset Reviews Last processed: 1/22/20, 3:25 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 recovered_host 344 336 0 4 1 4 5,850,651 1 1000 / 979 1 Complete
Data Ranges Review Images

Test Notes

  1. fail: data found in another stream (gaps: [1, 3] days: [1, 3])

Data Coverage

Deployment: 1
98%

Deployment Map

System Annotations

Metadata Start Date End Date Comment
GS05MOAS-GL560
12/20/15, 2:45 PM 11/22/16, 10:14 AM

Deployment 1: Glider was pushed far north of the Southern Ocean array where it was recovered. At recovery, barnacles were present mostly on the underside of the glider, and were not in an area to hinder flight performance.

Id: 578 By: lgarzio

Review Notes

Metadata Start Date End Date Comment
GS05MOAS

Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue.

By Lori Garzio, on 5/31/19

GS05MOAS-GL560-01-FLORDM000

Comparison with the shipboard CTD cast was done for D0001. Chl-a recorded by the glider was generally comparable to the CTD cast, with the exception of the chl-a maximum (~25 - 50 dbar) where the glider recorded higher chl-a values. Further analysis using the shipboard bottle data should be conducted. This is outside the scope of this project.

By Lori Garzio, on 3/11/19

New Note