Dep. | Preferred Method | Stream | DD | FD | SG | EG | Gaps | GD | TS | Rate (s) | Pressure Comp. | Time Order | Valid Data | Missing Data | Data Comp. | Missing Coords. | Review |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
1 | recovered_host | flort_o_glider_data | 298 | 242 | 0 | 56 | 1 | 1 | 900,417 | 1 | 200 / 209 | 1 | 1 | Complete |
Deployment: | 1 |
---|---|
flort_o_glider_data | 81% |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
GS05MOAS-PG566 |
12/20/15, 12:28 PM | 8/16/16, 8:40 PM |
Deployment 1: A leak developed in the external air bladder. Science sensors were turned off on 8/18/2016 and glider went into a low power drift state to preserve battery. Recovered in the Drake Passage with two broken wings and no rudder. Id: 579 By: lgarzio |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
GS05MOAS |
Glider optodes typically exhibit a sensor lag of 20-25 seconds. Example attached of several segments with consecutive down and up casts for CP05MOAS-GL335-04-DOSTAM000 on 10/26/2015 showing a typical measurement time lag. Profiles should be analyzed for individual time lags and data should be adjusted. OOI doesn't necessarily need to make the adjustment but users should be aware of the issue. By Lori Garzio, on 5/31/19 |
||
GS05MOAS-PG566-04-FLORTO000 |
There are no global ranges in the system for sci_bb3slo_b470_units, flort_o_bback_b470_total, flort_o_bback_b532_total, sci_bb3slo_b660_units, and flort_o_bback_b660_total. By Lori Garzio, on 2/3/20 |
||
GS05MOAS-PG566-04-FLORTO000 Deployment: 1 |
The arrays for the following variables are all NaNs: sci_bb3slo_b470_units, flort_o_bback_b470_total, flort_o_bback_b532_total, sci_bb3slo_b660_units, flort_o_bback_b660_total. According to the provenance.json (attached), the following cal values are missing: CC_1_dark_counts_volume_scatter, CC_1_scale_factor_volume_scatter, CC_1_measurement_wavelength, CC_2_dark_counts_volume_scatter, CC_2_scale_factor_volume_scatter, CC_2_measurement_wavelength, CC_3_dark_counts_volume_scatter, CC_3_scale_factor_volume_scatter, and CC_3_measurement_wavelength. By Lori Garzio, on 2/3/20 |