Dep. | Preferred Method | Stream | DD | FD | SG | EG | Gaps | GD | TS | Rate (s) | Pressure Comp. | Time Order | Valid Data | Missing Data | Data Comp. | Missing Coords. | Review |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
1 | recovered_inst | / | Complete | ||||||||||||||
2 | recovered_inst | 385 | 18 | 0 | 296 | 4 | 71 | 507,312 | 1 | / 1857 | 1 | Complete | |||||
3 | recovered_inst | / | Complete |
Deployment: | 1 | 2 | 3 |
---|---|---|---|
5% |
Deployment: | 1 | 2 | 3 |
---|---|---|---|
2 | 0.00 |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
RS01SBPD-DP01A-06-DOSTAD104 Method: recovered-inst Stream: dpc_optode_instrument_recovered |
8/8/16, 8:00 PM | 8/8/17, 8:00 PM |
The derived O2 products (L1 Temperature Compensated and L2 Salinity Corrected Dissolved Oxygen) are suspect, due to errors during vendor calibration. Until a correction can be applied, they should not be used from 08/09/2016 onwards. Id: 18 |
RS01SBPD-DP01A-06-DOSTAD104 Method: recovered-inst Stream: dpc_optode_instrument_recovered |
8/8/16, 8:00 PM | 8/8/17, 8:00 PM |
The derived O2 products (L1 Temperature Compensated and L2 Salinity Corrected Dissolved Oxygen) are suspect, due to errors during vendor calibration, and should not be used from 08/09/2016 onwards, until a correction can be applied. Id: 19 |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
RS01SBPD-DP01A |
There are no deployment depths for this profiler in asset management. By Lori Garzio, on 5/30/19 |
||
RS01SBPD-DP01A |
There are no data from deployments 1 and 3 available for download for any instrument. These data need to be ingested or should be annotated. By Lori Garzio, on 5/30/19 |
||
RS01SBPD-DP01A |
There is a limited amount of data from this profiler available for download. This should be annotated so users are aware of the issues with the deep profilers. By Lori Garzio, on 5/30/19 |
||
RS01SBPD-DP01A Deployment: 2 |