3-D Single Point Velocity Meter

Reference Designator
RS01SUM1-LJ01B-12-VEL3DB104
Review Status
Review Complete
Note
Depth
779m
Class
VEL3D (3-D Single Point Velocity Meter)
Make / Model
Nobska / MAVS-4

Dataset Reviews Last processed: 8/22/19, 3:52 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 streamed vel3d_b_sample 684 667 9 0 5 9 56,608,303 1 775 / 1 2 2 3 Complete
2 streamed vel3d_b_sample 384 379 1 0 3 4 32,418,718 775 / 1 2 2 3 Complete
3 streamed vel3d_b_sample 320 316 0 0 3 4 26,876,108 1 775 / 1 2 2 3 Complete
Data Ranges Review Images

Test Notes

  1. pass unique test - ascending not tested
  2. no other streams for comparison
  3. missing: ['pressure']

Data Coverage

Deployment: 123
vel3d_b_sample 98%99%99%

Lat/Lon Differences (km)

Deployment: 123
1 0.00
2 0.000.00
3 0.000.000.00

Example Composite Data Plot

System Annotations

Metadata Start Date End Date Comment
RS01SUM1
12/7/14, 2:45 PM 12/15/14, 7:00 PM

Note: PFE down. HVPS1 MOV explosion, 800A breaker tripped, investigation and restoration

Id: 1065 By: knuth

RS01SUM1
1/7/15, 2:32 AM 1/7/15, 3:06 AM

Note: PNWGP Portland <-> Seattle outage

Id: 1072 By: knuth

RS01SUM1
1/30/15, 7:00 PM 2/3/15, 7:00 PM

Note: Intermittent partial data loss due to storage drive problems at OTB

Id: 1079 By: knuth

RS01SUM1
3/21/15, 10:10 AM 3/22/15, 12:20 AM

Note: PNWGP outage due to City of Seattle fiber cable work

Id: 1086 By: knuth

RS01SUM1
6/12/15, 8:00 PM 6/15/15, 12:30 PM

Note: Network issues due to fire that damaged fibers between Portland and Seattle

Id: 1093 By: knuth

RS01SUM1
8/28/15, 8:00 PM 8/28/15, 8:30 PM

Note: Outage during major utility power failure in Seattle

Id: 1100 By: knuth

RS01SUM1
1/7/16, 1:10 AM 1/7/16, 1:52 AM

Note: Four 1-minute outages between Portland and Seattle due to maintenance

Id: 1107 By: knuth

RS01SUM1
3/10/16, 6:06 PM 3/11/16, 4:30 AM

Note: Fiber break between Portland and Seattle

Id: 1114 By: knuth

RS01SUM1
5/20/16, 12:33 PM 5/20/16, 2:04 PM

Note: Fiber break between Portland and Pacific City

Id: 1121 By: knuth

RS01SUM1
7/11/16, 10:53 PM 7/11/16, 11:51 PM

Note: Unexplained loss of power at Pittock Building in Portland

Id: 1128 By: knuth

RS01SUM1-LJ01B-12-VEL3DB104
Method: streamed
Stream: vel3d_b_sample
8/17/16, 5:52 PM 8/1/17, 11:00 AM

Two of 4 beams are reporting bad measurements. The instrument was bumped on frame during deployment, and may have been compromised. Investigation and resolution may have to wait until the next deployment in Summer 2017.

Id: 84 By: m.vardaro

RS01SUM1
12/17/16, 1:00 PM 12/17/16, 2:00 PM

Note: Corvalis data center lost power

Id: 1135 By: knuth

RS01SUM1
12/21/16, 8:50 PM 12/23/16, 7:44 AM

Note: Fiber break in Portland due to train crash

Id: 1142 By: knuth

RS01SUM1
1/8/17, 2:58 PM 1/8/17, 4:41 PM

Note: Network outage during major Seattle utility power failure

Id: 1149 By: knuth

RS01SUM1
1/9/17, 1:32 PM 1/9/17, 4:30 PM

Note: Lightning strike in Pacific City led to data interruption through isolation of both cable lines from shore station equipment.

Id: 1156 By: knuth

RS01SUM1
2/7/17, 8:00 AM 2/7/17, 10:00 AM

Note: Outage during PNWGP 1-hour router-maintenance

Id: 1163 By: knuth

RS01SUM1
2/15/17, 9:43 AM 2/15/17, 12:16 PM

Note: On Wednesday, February 15, power to the North and South cable lines of the Cabled Array was interrupted and all deployed infrastructure went offline. Power was restored on February 16. No hardware issue subsea was identified.

Id: 1170 By: knuth

RS01SUM1
1/18/18, 10:00 PM 1/19/18, 2:00 PM

A brief violent storm caused loss of power to both the North and South cable lines and the deployed infrastructure during this time. For more information please contact help@oceanobservatories.org. (Redmine # 13109)

Id: 331 By: friedrich.knuth
Flag: not_operational Exclude: No

RS01SUM1
8/22/18, 1:00 PM 8/24/18, 4:00 PM

On Wednesday Aug 22 2018, an automatic protective shutdown of primary power to the OOI Cabled Array occurred during installation of a new chiller system. Power was restored on Aug 24. Due to loss of power to the Cabled Array marine infrastructure, data were not collected by the instruments.

Id: 1488 By: lgarzio
Flag: not_operational Exclude: No

Review Notes

Metadata Start Date End Date Comment
RS01SUM1-LJ01B-12-VEL3DB104

The .nc files are missing the pressure coordinate.

By Lori Garzio, on 8/26/19

RS01SUM1-LJ01B-12-VEL3DB104

3% of seawater velocities are outside of global ranges. For eastward_turbulent_velocity there are periodic readings of >12 m/s, northward_turbulent_velocity periodic readings of 7 m/s, and upward_turbulent_velocity periodic readings of 10 m/s.

By Lori Garzio, on 8/26/19

RS01SUM1-LJ01B-12-VEL3DB104
Deployment: 2

No valid seawater velocity data for this deployment - see annotation ID 84

By Lori Garzio, on 8/26/19

RS01SUM1-LJ01B-12-VEL3DB104
Deployment: 3
8/11/17, 11:00 PM 6/26/18, 3:30 AM

Seawater velocities look suspect for this deployment and don't follow the same pattern as the velocities for deployment 1. This should be investigated and the dataset should be annotated if there was an issue.

By Lori Garzio, on 8/26/19

New Note