Hydrothermal Vent Fluid Temperature and Resistivity

Reference Designator
RS03INT1-MJ03C-10-TRHPHA301
Review Status
Review Complete
Note
Depth
1,520m
Class
TRHPH (Hydrothermal Vent Fluid Temperature and Resistivity)
Make / Model
Lilley, UW / Vent Temperature & Resistivity

Dataset Reviews Last processed: 8/20/19, 12:50 PM

QC Check Info
Dep. Preferred Method Stream DD FD SG EG Gaps GD TS Rate (s) Pressure Comp. Time Order Valid Data Missing Data Data Comp. Missing Coords. Review
1 streamed trhph_sample 287 278 0 0 6 11 1,166,739 19 1520 / 1 1 2 Complete
2 streamed trhph_sample 382 378 0 0 1 4 1,604,281 19 1520 / 1 1 2 Complete
3 streamed trhph_sample 384 384 0 0 1 1 1,647,534 19 1520 / 1 1 2 Complete
Data Ranges Review Images

Test Notes

  1. no other streams for comparison
  2. missing: ['pressure']

Data Coverage

Deployment: 123
trhph_sample 97%99%100%

Lat/Lon Differences (km)

Deployment: 123
1 0.00
2 0.000.00
3 0.000.000.00

System Annotations

Metadata Start Date End Date Comment
RS03INT1
12/7/14, 2:45 PM 12/15/14, 7:00 PM

Note: PFE down. HVPS1 MOV explosion, 800A breaker tripped, investigation and restoration

Id: 1070 By: knuth

RS03INT1
1/7/15, 2:32 AM 1/7/15, 3:06 AM

Note: PNWGP Portland <-> Seattle outage

Id: 1077 By: knuth

RS03INT1
1/30/15, 7:00 PM 2/3/15, 7:00 PM

Note: Intermittent partial data loss due to storage drive problems at OTB

Id: 1084 By: knuth

RS03INT1
3/21/15, 10:10 AM 3/22/15, 12:20 AM

Note: PNWGP outage due to City of Seattle fiber cable work

Id: 1091 By: knuth

RS03INT1
6/12/15, 8:00 PM 6/15/15, 12:30 PM

Note: Network issues due to fire that damaged fibers between Portland and Seattle

Id: 1098 By: knuth

RS03INT1
8/28/15, 8:00 PM 8/28/15, 8:30 PM

Note: Outage during major utility power failure in Seattle

Id: 1105 By: knuth

RS03INT1
1/7/16, 1:10 AM 1/7/16, 1:52 AM

Note: Four 1-minute outages between Portland and Seattle due to maintenance

Id: 1112 By: knuth

RS03INT1
3/10/16, 6:06 PM 3/11/16, 4:30 AM

Note: Fiber break between Portland and Seattle

Id: 1119 By: knuth

RS03INT1
5/20/16, 12:33 PM 5/20/16, 2:04 PM

Note: Fiber break between Portland and Pacific City

Id: 1126 By: knuth

RS03INT1
7/11/16, 10:53 PM 7/11/16, 11:51 PM

Note: Unexplained loss of power at Pittock Building in Portland

Id: 1133 By: knuth

RS03INT1
12/17/16, 1:00 PM 12/17/16, 2:00 PM

Note: Corvalis data center lost power

Id: 1140 By: knuth

RS03INT1
12/21/16, 8:50 PM 12/23/16, 7:44 AM

Note: Fiber break in Portland due to train crash

Id: 1147 By: knuth

RS03INT1
1/8/17, 2:58 PM 1/8/17, 4:41 PM

Note: Network outage during major Seattle utility power failure

Id: 1154 By: knuth

RS03INT1
1/9/17, 1:32 PM 1/9/17, 4:30 PM

Note: Lightning strike in Pacific City led to data interruption through isolation of both cable lines from shore station equipment.

Id: 1161 By: knuth

RS03INT1
2/7/17, 8:00 AM 2/7/17, 10:00 AM

Note: Outage during PNWGP 1-hour router-maintenance

Id: 1168 By: knuth

RS03INT1
2/15/17, 9:43 AM 2/15/17, 12:16 PM

Note: On Wednesday, February 15, power to the North and South cable lines of the Cabled Array was interrupted and all deployed infrastructure went offline. Power was restored on February 16. No hardware issue subsea was identified.

Id: 1175 By: knuth

RS03INT1
1/18/18, 10:00 PM 1/19/18, 2:00 PM

A brief violent storm caused loss of power to both the North and South cable lines and the deployed infrastructure during this time. For more information please contact help@oceanobservatories.org. (Redmine # 13109)

Id: 324 By: friedrich.knuth
Flag: not_operational Exclude: No

RS03INT1
8/22/18, 1:00 PM 8/24/18, 4:00 PM

On Wednesday Aug 22 2018, an automatic protective shutdown of primary power to the OOI Cabled Array occurred during installation of a new chiller system. Power was restored on Aug 24. Due to loss of power to the Cabled Array marine infrastructure, data were not collected by the instruments.

Id: 1495 By: lgarzio
Flag: not_operational Exclude: No

Review Notes

Metadata Start Date End Date Comment
RS03INT1-MJ03C-10-TRHPHA301

There is no pressure variable in any of the .nc files. This should be a coordinate.

By Lori Garzio, on 9/23/19

RS03INT1-MJ03C-10-TRHPHA301

There are no global ranges in the system for resistivity_temp_degc, vent_fluid_orp and trhph_thermistor_temp

By Lori Garzio, on 9/23/19

RS03INT1-MJ03C-10-TRHPHA301
Deployment: 1
9/27/14, 12:00 AM 7/10/15, 12:00 AM

Suspect data for deployment 1 - vent_fluid_temperaure drops off to zero. This should be annotated.

By Lori Garzio, on 9/23/19

RS03INT1-MJ03C-10-TRHPHA301
Deployment: 2

72% of vent_fluid_chloride_conc values are NaNs. Values spike at the end of Oct 2015 and stop.

By Lori Garzio, on 9/23/19

RS03INT1-MJ03C-10-TRHPHA301
Deployment: 2

10% of vent_fluid_orp values are outliers

By Lori Garzio, on 9/23/19

RS03INT1-MJ03C-10-TRHPHA301
Deployment: 3

vent_fluid_orp values are suspect

By Lori Garzio, on 9/23/19

RS03INT1-MJ03C-10-TRHPHA301
Deployment: 3

72% of vent_fluid_chloride_conc values are NaNs. Values spike in Nov 2016 and stop.

By Lori Garzio, on 9/23/19

New Note