Method | Data Stream | Content | Type | |
---|---|---|---|---|
recovered_host | pco2w_abc_dcl_instrument_blank_recovered | Data Products | Report M2M Stats | Science |
recovered_host | pco2w_abc_dcl_instrument_recovered | Data Products | Report M2M Stats | Science |
recovered_host | pco2w_abc_dcl_metadata_recovered | Metadata | Report M2M | Engineering |
recovered_host | pco2w_abc_dcl_power_recovered | Power Data | Report M2M | Engineering |
recovered_inst | pco2w_abc_instrument | Data Products | Report M2M Stats | Science |
recovered_inst | pco2w_abc_instrument_blank | Data Products | Report M2M Stats | Science |
recovered_inst | pco2w_abc_metadata | Metadata | Report M2M | Engineering |
recovered_inst | pco2w_abc_power | Power Data | Report M2M | Engineering |
telemetered | pco2w_abc_dcl_instrument | Data Products | Report M2M Stats | Science |
telemetered | pco2w_abc_dcl_instrument_blank | Data Products | Report M2M Stats | Science |
telemetered | pco2w_abc_dcl_metadata | Metadata | Report M2M | Engineering |
telemetered | pco2w_abc_dcl_power | Power Data | Report M2M | Engineering |
Deployment | Cruise | Start Date | Stop Date | Mooring Asset | Node Asset | Sensor Asset | Latitude | Longitude | Deployment Depth | Water Depth |
---|---|---|---|---|---|---|---|---|---|---|
1 Review | OC1410A | 10/07/2014 | 04/10/2015 | CGMCE-06ISSM-00001 | CGINS-PCO2WB-C0062 | 47.1334 | -124.268 | 30 | 30 | |
2 Review | OC1503D | 04/10/2015 | 10/13/2015 | CGMCE-06ISSM-00002 | CGINS-PCO2WB-C0085 | 47.1311 | -124.272 | 29 | 29 | |
3 Review | OC1510A | 10/13/2015 | 05/06/2016 | CGMCE-06ISSM-00003 | CGINS-PCO2WB-C0062 | 47.1333 | -124.271 | 29 | 29 | |
4 Review | TN-342 | 05/06/2016 | 09/24/2016 | CGMCE-06ISSM-00004 | CGINS-PCO2WB-C0063 | 47.135 | -124.273 | 29 | 29 | |
5 Review | AT37-03 | 09/27/2016 | 04/15/2017 | CGMCE-06ISSM-00005 | CGINS-PCO2WB-C0081 | 47.1336 | -124.272 | 29 | 29 | |
6 Review | SKQ201704S | 04/13/2017 | 10/05/2017 | CGMCE-06ISSM-00006 | CGINS-PCO2WB-C0119 | 47.1341 | -124.272 | 29 | 29 | |
7 Review | SKQ201715S | 10/05/2017 | 03/28/2018 | CGMCE-06ISSM-00007 | CGINS-PCO2WB-C0063 | 47.1346 | -124.272 | 29 | 29 | |
8 Review | SKQ201808S | 03/28/2018 | 09/20/2018 | CGMCE-06ISSM-00008 | CGINS-PCO2WB-C0119 | 47.1336 | -124.269 | 29 | 29 | |
9 Review | SR1813 | 09/24/2018 | 04/25/2019 | CGMCE-06ISSM-00009 | CGINS-PCO2WB-C0084 | 47.1348 | -124.271 | 29 | 29 | |
10 Review | SKQ201910S | 04/25/2019 | 10/12/2019 | CGMCE-06ISSM-00010 | CGINS-PCO2WB-C0147 | 47.1347 | -124.271 | 29 | 29 | |
11 Review | SKQ201921S | 10/12/2019 | CGMCE-06ISSM-00011 | CGINS-PCO2WB-C0053 | 47.1344 | -124.271 | 29 | 29 |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
CE06ISSM-MFD35-05-PCO2WB000 |
There are no recovered_inst data available for download for any deployment. According to the ingest csvs, the recovered_inst raw data files are available for all deployments. These files should be ingested or the datasets should be annotated to explain why the recovered_inst data aren't available. By Lori Garzio, on 7/24/19 |
||
CE06ISSM-MFD35-05-PCO2WB000 |
There is no pressure variable in the .nc files. This should be a coordinate. By Lori Garzio, on 7/24/19 |
||
CE06ISSM-MFD35-05-PCO2WB000 |
Data generally look suspect for many deployments of this instrument because of the high percentage of values outside of global ranges. The dataset should be annotated if there is a known issue with this instrument to inform users if all or some data should be considered suspect. By Lori Garzio, on 7/24/19 |
||
CE06ISSM-MFD35-05-PCO2WB000 Deployment: 1 |
|||
CE06ISSM-MFD35-05-PCO2WB000 Deployment: 2 |
|||
CE06ISSM-MFD35-05-PCO2WB000 Deployment: 4 |
|||
CE06ISSM-MFD35-05-PCO2WB000 Deployment: 5 |
|||
CE06ISSM-MFD35-05-PCO2WB000 Deployment: 6 |
|||
CE06ISSM-MFD35-05-PCO2WB000 Deployment: 7 |
Metadata | Start Date | End Date | Comment |
---|---|---|---|
CE06ISSM-MFD35-05-PCO2WB000 Method: telemetered Stream: pco2w_abc_dcl_instrument |
10/7/14, 3:45 PM | 1/9/19, 7:00 PM |
UPDATED, 2019-01-15: After careful review, the original data product algorithm (pco2_pco2wat) for this sensor was found to contain an error that resulted in calculation of incorrect derived L1 pCO2 values. In consultation with the vendor, the algorithm has been corrected, verified, and uploaded to the OOI system. This update, effective 2019-01-10, applies to all deployments for this sensor. All derived L1 pCO2 data products generated prior to this date should be re-requested in order to ensure utilization of the correct algorithm. The updated algorithm can be viewed at https://github.com/oceanobservatories/ion-functions/blob/master/ion_functions/data/co2_functions.py. * Until further notice, derived L1 pCO2 data products from this sensor are incorrect and should not be used. A significant discrepancy has been uncovered between the pCO2 values from shipboard water samples vs. values calculated by OOI system, and the algorithms are under review: https://github.com/ooici/ion-functions/blob/master/ion_functions/data/co2_functions.py. To receive instant notification of updates to the code, add yourself as a watcher to the GitHub repository linked above. For further questions please email help@oceanobservatories.org. Id: 271 By: friedrich.knuth |
CE06ISSM-MFD35-05-PCO2WB000 Method: recovered_host Stream: pco2w_abc_dcl_instrument_recovered |
10/7/14, 3:45 PM | 1/9/19, 7:00 PM |
UPDATED, 2019-01-15: After careful review, the original data product algorithm (pco2_pco2wat) for this sensor was found to contain an error that resulted in calculation of incorrect derived L1 pCO2 values. In consultation with the vendor, the algorithm has been corrected, verified, and uploaded to the OOI system. This update, effective 2019-01-10, applies to all deployments for this sensor. All derived L1 pCO2 data products generated prior to this date should be re-requested in order to ensure utilization of the correct algorithm. The updated algorithm can be viewed at https://github.com/oceanobservatories/ion-functions/blob/master/ion_functions/data/co2_functions.py. * Until further notice, derived L1 pCO2 data products from this sensor are incorrect and should not be used. A significant discrepancy has been uncovered between the pCO2 values from shipboard water samples vs. values calculated by OOI system, and the algorithms are under review: https://github.com/ooici/ion-functions/blob/master/ion_functions/data/co2_functions.py. To receive instant notification of updates to the code, add yourself as a watcher to the GitHub repository linked above. For further questions please email help@oceanobservatories.org. Id: 295 By: friedrich.knuth |
CE06ISSM |
10/7/14, 3:45 PM | 9/30/16, 8:00 PM |
All of the inshore surface moorings suffer from an issue called the 'top of the day problem.' A race condition exists in the CPM firmware when transferring daily log files harvested from the various sub-components in a buoy system (e.g. DCL16, CPM3, DCL35, etc) over the Iridium RUDICS connection. The primary CPM rsync's data files from the various sub-components to a central data directory. The first time a daily file is created/copied into that directory (usually shortly after midnight), a compressed copy of the file is created and that compressed file is queued up in a separate text file for transmission to shore. The transfer routine reads that text file and processes the files from that list in order from top to bottom. Once that compressed file is transferred to shore, no further updates for that file are sent to shore. Meanwhile, the CPM continues to rsync data from the different sub-components, updating the original daily log file. The net result is that data is accumulating in the daily log file for an instrument, but we may see only one record (or none) depending on when the rsync job occurred and the compressed file was created and staged for transfer to shore. This affects telemetered data only. Id: 897 By: michaesm |
CE06ISSM |
7/26/16, 8:00 PM | 10/3/16, 8:00 PM |
Correcting the iridium data telemetry issues, permitting full telemetry of data files, resulting in greater power consumption than anticipated. Result is power level of the battery pack has dropped below operational limits and subsystems can no longer power instruments. Shutting the mooring down, and placing in a low power maintenance mode. 11 instruments offline.
Id: 898
By: michaesm
|
CE06ISSM |
12/31/16, 7:00 PM | 4/22/17, 8:00 PM |
The buoy main battery voltage is being drawn down faster than expected. Have shutdown further power to DCL17 and to the OPTAA on DCL16. This takes the following instruments offline: Id: 899 By: michaesm |
CE06ISSM-MFD35 |
5/10/17, 8:00 PM | 10/30/17, 8:00 PM |
Lost communication with DCL35, on the MFN of CE06ISSM sometime after 2017/05/11 12:32:17.552. Suspect failed/corrupted SD card based on results of minicom session with DCL35 (see attached). With the exception of the VEL3D, all instruments hosted by DCL35 are autonomous. No further data should be expected from: ADCPT (recording internally), PRESF (recording internally), PC02W (recording internally), PHSEN (recording internally), VEL3D (offline)
Id: 900
By: michaesm
|
CE06ISSM |
10/2/17, 8:00 PM | 3/2/18, 3:44 AM |
CAMDS not serviced by vendor in time for deployment
Id: 1236
By: michaesm
|
CE06ISSM-MFD35 Method: telemetered |
10/17/19, 12:45 AM | 5/23/20, 8:00 PM |
Updated, 2020-05-24: Between 2019-10-17 and 2020-05-24, the clock on CPM3 was off by approximately 27 minutes. Communications between the buoy and the MFN were out of sync during this time period. The clock was automatically updated during routine operations on May 24, 2020 to the correct time, allowing communications and normal operations on the MFN to resume. *Communications to the seafloor sensors failed. Status of the instrumentation and data loggers is unknown. With the exception of the VEL3D and OPTAA, the sensors on the seafloor platform should continue to autonomously sample for the remainder of deployment. Until we recover and redeploy the mooring in the spring of 2020, no further telemetered data will be available from the seafloor sensors.
Id: 1758
By: wingardc
|