3-Wavelength Fluorometer

Reference Designator
CE06ISSM-RID16-02-FLORTD000
Start Depth
7
End Depth
7
Location
Near Surface Instrument Frame
Current Status
Review Complete
Class
FLORT (3-Wavelength Fluorometer)
Series
FLORT-D
Science Discipline
Biological
Make
WET Labs
Model
ECO Triplet-w
M2M Example
https://ooinet.oceanobservatories.org/api/m2m/12576/sensor/inv/CE06ISSM/RID16/02-FLORTD000/metadata
Method Data Stream Content Type
recovered_host flort_sample Data Products Report M2M Stats Science
telemetered flort_sample Data Products Report M2M Stats Science
Deployment Cruise Start Date Stop Date Mooring Asset Node Asset Sensor Asset Latitude Longitude Deployment Depth Water Depth
1 Review OC1410A 10/07/2014 04/10/2015 CGMCE-06ISSM-00001 CGINS-FLORTD-00996 47.1334 -124.268 7 30
2 Review OC1503D 04/10/2015 10/13/2015 CGMCE-06ISSM-00002 CGINS-FLORTD-01197 47.1311 -124.272 7 29
3 Review OC1510A 10/13/2015 05/06/2016 CGMCE-06ISSM-00003 CGINS-FLORTD-01121 47.1333 -124.271 7 29
4 Review TN-342 05/06/2016 09/24/2016 CGMCE-06ISSM-00004 CGINS-FLORTD-01197 47.135 -124.273 7 29
5 Review AT37-03 09/27/2016 04/15/2017 CGMCE-06ISSM-00005 CGINS-FLORTD-01302 47.1336 -124.272 7 29
6 Review SKQ201704S 04/13/2017 10/05/2017 CGMCE-06ISSM-00006 CGINS-FLORTD-01152 47.1341 -124.272 7 29
7 Review SKQ201715S 10/05/2017 03/28/2018 CGMCE-06ISSM-00007 CGINS-FLORTD-01151 47.1346 -124.272 7 29
8 Review SKQ201808S 03/28/2018 09/20/2018 CGMCE-06ISSM-00008 CGINS-FLORTD-01155 47.1336 -124.269 7 29
9 Review SR1813 09/24/2018 04/25/2019 CGMCE-06ISSM-00009 CGINS-FLORTD-01151 47.1348 -124.271 7 29
10 Review SKQ201910S 04/25/2019 10/12/2019 CGMCE-06ISSM-00010 CGINS-FLORTD-01488 47.1347 -124.271 7 29
11 Review SKQ201921S 10/12/2019 CGMCE-06ISSM-00011 CGINS-FLORTD-00996 47.1344 -124.271 7 29
Metadata Start Date End Date Comment
CE06ISSM-RID16-02-FLORTD000

pressure_depth is an array of fill values for every deployment. This variable should contain valid pressure data.

By Lori Garzio, on 6/28/19

CE06ISSM-RID16-02-FLORTD000
Deployment: 1

Annotation IDs 97 and 98 should be updated: "Volume scattering count data between October 2014 and April 2015 show a repetitive pattern of steady increase in signal to the upper detection limit of the instrument, followed by a drop back to zero. Investigation is ongoing."

By Lori Garzio, on 7/2/19

CE06ISSM-RID16-02-FLORTD000
Deployment: 3

A substantial amount of CDOM values are negative

By Lori Garzio, on 7/2/19

CE06ISSM-RID16-02-FLORTD000
Deployment: 4
5/6/16, 2:24 AM 5/25/16, 12:00 AM

Suspect noise in the CDOM data at the beginning of the deployment.

By Lori Garzio, on 7/2/19

CE06ISSM-RID16-02-FLORTD000
Deployment: 6
6/15/17, 12:00 AM 10/5/17, 1:53 PM

Suspect data - potential biofouling

By Lori Garzio, on 7/2/19

New Note

Metadata Start Date End Date Comment
CE06ISSM
10/7/14, 3:45 PM 9/30/16, 8:00 PM

All of the inshore surface moorings suffer from an issue called the 'top of the day problem.' A race condition exists in the CPM firmware when transferring daily log files harvested from the various sub-components in a buoy system (e.g. DCL16, CPM3, DCL35, etc) over the Iridium RUDICS connection. The primary CPM rsync's data files from the various sub-components to a central data directory. The first time a daily file is created/copied into that directory (usually shortly after midnight), a compressed copy of the file is created and that compressed file is queued up in a separate text file for transmission to shore. The transfer routine reads that text file and processes the files from that list in order from top to bottom. Once that compressed file is transferred to shore, no further updates for that file are sent to shore. Meanwhile, the CPM continues to rsync data from the different sub-components, updating the original daily log file. The net result is that data is accumulating in the daily log file for an instrument, but we may see only one record (or none) depending on when the rsync job occurred and the compressed file was created and staged for transfer to shore. This affects telemetered data only.

Id: 897 By: michaesm

CE06ISSM-RID16-02-FLORTD000
Method: recovered_host
Stream: flort_dj_dcl_instrument_recovered
10/12/14, 3:45 PM 4/9/15, 9:33 PM

Volume scattering count data between October 2014 and April 2015 show a repetitive pattern of steady increase in signal to the upper detection limit of the instrument, followed by a drop back to zero. Investigation is ongoing.

Id: 97 By: m.vardaro
Flag: suspect Exclude: No

CE06ISSM-RID16-02-FLORTD000
Method: telemetered
Stream: flort_dj_dcl_instrument
10/13/14, 8:30 PM 4/10/15, 7:33 AM

Volume scattering count data between October 2014 and April 2015 show a repetitive pattern of steady increase in signal to the upper detection limit of the instrument, followed by a drop back to zero. Investigation is ongoing.

Id: 98 By: m.vardaro
Flag: suspect Exclude: No

CE06ISSM
7/26/16, 8:00 PM 10/3/16, 8:00 PM

Correcting the iridium data telemetry issues, permitting full telemetry of data files, resulting in greater power consumption than anticipated. Result is power level of the battery pack has dropped below operational limits and subsystems can no longer power instruments. Shutting the mooring down, and placing in a low power maintenance mode. 11 instruments offline.

Id: 898 By: michaesm
Flag: not_operational Exclude: No

CE06ISSM
12/31/16, 7:00 PM 4/22/17, 8:00 PM

The buoy main battery voltage is being drawn down faster than expected. Have shutdown further power to DCL17 and to the OPTAA on DCL16. This takes the following instruments offline:
DCL17
MOPAK
CTDBP (with attached FLORT)
VELPT
ACOMM
OPTAA
Have had to power off DCL17 and DCL16 and reduce call in frequency to 12 hours. Will still (at least should) receive data from MFN. Instruments affected are:
DCL17 (Buoy)
MOPAK
CTDBP3 (with FLORT)
VELPT1
UCSPP (aka ACOMM)
DCL16 (NSIF)
CTDBP1 (with DOSTA) (should internally log)
FLORT
PCO2W1 (should internally log)
PHSEN1 (should internally log)
OPTAA1
NUTNR (should internally log)
SPKIR
VELPT2 (should internally log)

Id: 899 By: michaesm

CE06ISSM
10/2/17, 8:00 PM 3/2/18, 3:44 AM

CAMDS not serviced by vendor in time for deployment

Id: 1236 By: michaesm
Flag: not_operational Exclude: No

CE06ISSM-RID16-02-FLORTD000
Method: telemetered
5/23/20, 8:00 PM 6/22/20, 2:45 PM

Reduced battery power on the mooring (unforeseen delay of mooring recovery well past the originally scheduled date) has resulted in the decision to reduce the operational schedule for this sensor for. Data, at a reduced sample rate, will be collected and telemetered to shore for as long as possible.

Id: 2197 By: wingardc

CE06ISSM-RID16-02-FLORTD000
Method: telemetered
6/22/20, 2:45 PM

Reduced battery power on the mooring (unforeseen delay of mooring recovery well past the originally scheduled date) has resulted in the decision to power off this sensor for the remainder of the deployment.

Id: 2200 By: wingardc
Flag: not_operational Exclude: No