3-Axis Motion Pack

Reference Designator
CE06ISSM-SBD17-01-MOPAK0000
Start Depth
0
End Depth
0
Location
Buoy Well, center of Mass
Current Status
Engineering
Class
MOPAK (3-Axis Motion Pack)
Series
MOPAK-0
Science Discipline
Physical
Make
MicroStrain
Model
3DM-GX3-25
M2M Example
https://ooinet.oceanobservatories.org/api/m2m/12576/sensor/inv/CE06ISSM/SBD17/01-MOPAK0000/metadata
Method Data Stream Content Type
recovered_host mopak_o_dcl_accel_recovered Accelerometer Data Report M2M Engineering
recovered_host mopak_o_dcl_rate_recovered Rate Data Report M2M Engineering
telemetered mopak_o_dcl_accel Accelerometer Data Report M2M Engineering
telemetered mopak_o_dcl_rate Rate Data Report M2M Engineering
Deployment Cruise Start Date Stop Date Mooring Asset Node Asset Sensor Asset Latitude Longitude Deployment Depth Water Depth
1 Review OC1410A 10/07/2014 04/10/2015 CGMCE-06ISSM-00001 R00007 47.1334 -124.268 0 30
2 Review OC1503D 04/10/2015 10/13/2015 CGMCE-06ISSM-00002 OL000214 47.1311 -124.272 0 29
3 Review OC1510A 10/13/2015 05/06/2016 CGMCE-06ISSM-00003 OL000215 47.1333 -124.271 0 29
4 Review TN-342 05/06/2016 09/24/2016 CGMCE-06ISSM-00004 OL000398 47.135 -124.273 0 29
5 Review AT37-03 09/27/2016 04/15/2017 CGMCE-06ISSM-00005 CGINS-MOPAK0-11923 47.1336 -124.272 0 29
6 Review SKQ201704S 04/13/2017 10/05/2017 CGMCE-06ISSM-00006 CGINS-MOPAK0-11845 47.1341 -124.272 0 29
7 Review SKQ201715S 10/05/2017 03/28/2018 CGMCE-06ISSM-00007 CGINS-MOPAK0-11923 47.1346 -124.272 0 29
8 Review SKQ201808S 03/28/2018 09/20/2018 CGMCE-06ISSM-00008 CGINS-MOPAK0-11845 47.1336 -124.269 0 29
9 Review SR1813 09/24/2018 04/25/2019 CGMCE-06ISSM-00009 CGINS-MOPAK0-11923 47.1348 -124.271 0 29
10 Review SKQ201910S 04/25/2019 10/12/2019 CGMCE-06ISSM-00010 CGINS-MOPAK0-11924 47.1347 -124.271 0 29
11 Review SKQ201921S 10/12/2019 CGMCE-06ISSM-00011 CGINS-MOPAK0-11923 47.1344 -124.271 0 29

No notes yet.

New Note

Metadata Start Date End Date Comment
CE06ISSM
10/7/14, 3:45 PM 9/30/16, 8:00 PM

All of the inshore surface moorings suffer from an issue called the 'top of the day problem.' A race condition exists in the CPM firmware when transferring daily log files harvested from the various sub-components in a buoy system (e.g. DCL16, CPM3, DCL35, etc) over the Iridium RUDICS connection. The primary CPM rsync's data files from the various sub-components to a central data directory. The first time a daily file is created/copied into that directory (usually shortly after midnight), a compressed copy of the file is created and that compressed file is queued up in a separate text file for transmission to shore. The transfer routine reads that text file and processes the files from that list in order from top to bottom. Once that compressed file is transferred to shore, no further updates for that file are sent to shore. Meanwhile, the CPM continues to rsync data from the different sub-components, updating the original daily log file. The net result is that data is accumulating in the daily log file for an instrument, but we may see only one record (or none) depending on when the rsync job occurred and the compressed file was created and staged for transfer to shore. This affects telemetered data only.

Id: 897 By: michaesm

CE06ISSM
7/26/16, 8:00 PM 10/3/16, 8:00 PM

Correcting the iridium data telemetry issues, permitting full telemetry of data files, resulting in greater power consumption than anticipated. Result is power level of the battery pack has dropped below operational limits and subsystems can no longer power instruments. Shutting the mooring down, and placing in a low power maintenance mode. 11 instruments offline.

Id: 898 By: michaesm
Flag: not_operational Exclude: No

CE06ISSM
12/31/16, 7:00 PM 4/22/17, 8:00 PM

The buoy main battery voltage is being drawn down faster than expected. Have shutdown further power to DCL17 and to the OPTAA on DCL16. This takes the following instruments offline:
DCL17
MOPAK
CTDBP (with attached FLORT)
VELPT
ACOMM
OPTAA
Have had to power off DCL17 and DCL16 and reduce call in frequency to 12 hours. Will still (at least should) receive data from MFN. Instruments affected are:
DCL17 (Buoy)
MOPAK
CTDBP3 (with FLORT)
VELPT1
UCSPP (aka ACOMM)
DCL16 (NSIF)
CTDBP1 (with DOSTA) (should internally log)
FLORT
PCO2W1 (should internally log)
PHSEN1 (should internally log)
OPTAA1
NUTNR (should internally log)
SPKIR
VELPT2 (should internally log)

Id: 899 By: michaesm

CE06ISSM
10/2/17, 8:00 PM 3/2/18, 3:44 AM

CAMDS not serviced by vendor in time for deployment

Id: 1236 By: michaesm
Flag: not_operational Exclude: No

CE06ISSM-SBD17-01-MOPAK0000
Method: telemetered
5/18/20, 8:35 PM

Reduced battery power on the mooring (unforeseen delay of mooring recovery well past the originally scheduled date) has resulted in the decision to power off this sensor for the remainder of the deployment.

Id: 2187 By: wingardc
Flag: not_operational Exclude: No