Spectrophotometer

Reference Designator
CE07SHSM-MFD37-01-OPTAAD000
Start Depth
87
End Depth
87
Location
Multi-Function Node
Current Status
Review Complete
Class
OPTAA (Absorption Spectrophotometer)
Series
OPTAA-D
Science Discipline
Biological
Make
WET Labs
Model
AC-S
M2M Example
https://ooinet.oceanobservatories.org/api/m2m/12576/sensor/inv/CE07SHSM/MFD37/01-OPTAAD000/metadata
Method Data Stream Content Type
recovered_host optaa_dj_dcl_instrument_recovered Data Products Report M2M Stats Science
recovered_host optaa_dj_dcl_metadata_recovered Metadata Report M2M Engineering
telemetered optaa_dj_dcl_instrument Data Products Report M2M Stats Science
telemetered optaa_dj_dcl_metadata Metadata Report M2M Engineering
Deployment Cruise Start Date Stop Date Mooring Asset Node Asset Sensor Asset Latitude Longitude Deployment Depth Water Depth
1 Review OC1503D 04/05/2015 09/23/2015 CGMCE-07SHSM-00001 CGINS-OPTAAD-00127 46.9861 -124.566 87 87
2 Review OC1510A 10/13/2015 05/13/2016 CGMCE-07SHSM-00002 CGINS-OPTAAD-00182 46.9875 -124.569 87 87
3 Review TN-342 05/05/2016 09/27/2016 CGMCE-07SHSM-00003 CGINS-OPTAAD-00149 46.9895 -124.569 87 87
5 Review SKQ201704S 04/11/2017 10/06/2017 CGMCE-07SHSM-00005 CGINS-OPTAAD-00135 46.9882 -124.567 89 89
8 Review SR1813 09/18/2018 04/26/2019 CGMCE-07SHSM-00008 CGINS-OPTAAD-00261 46.9835 -124.564 87 87
9 Review SKQ201910S 04/24/2019 10/12/2019 CGMCE-07SHSM-00009 CGINS-OPTAAD-00184 46.9878 -124.568 87 87
10 Review SKQ201921S 10/10/2019 CGMCE-07SHSM-00010 CGINS-OPTAAD-00179 46.9841 -124.566 87 87
Metadata Start Date End Date Comment
CE07SHSM-MFD37-01-OPTAAD000

According to Roesler and Barnard (Optical proxy for phytoplankton biomass in the absence of photophysiology: Rethinking the absorption line height, Methods in Oceanography, 2013), “absorption meters are highly prone to biofouling, particularly biofilms which not only attenuate the collimated beam but also impact the scattering properties of the optical surfaces and tubes. In productive coastal waters biofouling can have significant impacts (i.e. 10% of the signal) within one to two weeks”.

By Lori Garzio, on 6/28/19

CE07SHSM-MFD37-01-OPTAAD000

There is no pressure variable in the .nc files. This should be a coordinate.

By Lori Garzio, on 6/28/19

CE07SHSM-MFD37-01-OPTAAD000
Deployment: 2

There are no data available for download for this deployment. According to the ingest csv, "Communications with MFN failed immediately upon deployment, no data will be available from this instrument". This should be annotated in the system.

By Lori Garzio, on 6/28/19

New Note

Metadata Start Date End Date Comment
CE07SHSM-MFD37-01-OPTAAD000
4/5/15, 1:15 PM

Per vendor feedback, the first 1-2 min of each OPTAA instrument sampling period should be considered suspect. Signal drift can occur as the instrument warms up, flushes with seawater, and the filter wheel spins up. Users are encouraged to delete the first minute of data, at a minimum, from each measurement burst and then calculate a median average of the remaining data from the burst before working further with the OPTAA data.

Id: 2099 By: wingardc

CE07SHSM-MFD37
8/11/16, 8:00 PM 9/26/16, 8:00 PM

The Power System Controller on CE07SHSM indicates the power connection to the MFN short-circuited on 2016-08-12 at 23:30 UTC. Failure is indicated by decreased voltage (70 VDC instead of 380 VDC) and increased current draw (1900 mA instead of 120 mA) to the CVT. This symptom is indicative of a short-circuit in the power lines of either the stretch hose (most likely) or EM Chain. In the past, this failure has been associated with a failed stretch hose. Have disabled all further power and communications with the MFN. This means no more telemetered data from: ADCPT2, VEL3D, CTDBP2 + DOSTA, PRESF, PCO2W, PHSEN2, OPTAA2, ZPLSC

Id: 917 By: michaesm
Flag: not_available Exclude: No

CE07SHSM-MFD37-01-OPTAAD000
8/11/16, 8:00 PM 9/26/16, 8:00 PM

The Power System Controller on CE07SHSM indicates the power connection to the MFN short-circuited on 2016-08-12 at 23:30 UTC. Failure is indicated by decreased voltage (70 VDC instead of 380 VDC) and increased current draw (1900 mA instead of 120 mA) to the CVT. This symptom is indicative of a short-circuit in the power lines of either the stretch hose (most likely) or EM Chain. In the past, this failure has been associated with a failed stretch hose. Have disabled all further power and communications with the MFN. Therefore, there is no telemetered data for this reference designator.

Id: 918 By: michaesm
Flag: not_available Exclude: No

CE07SHSM-MFD37-01-OPTAAD000
9/19/16, 8:00 PM 4/30/17, 8:00 PM

Instrument not deployed during Fall 2016 cruise because its refurbishment was not complete before the cruise.

Id: 919 By: michaesm
Flag: not_operational Exclude: No

CE07SHSM
12/1/17, 9:35 AM 3/25/18, 3:09 PM

Data loss due to a power system failure. The mooring stopped receiving power from photo voltaic panel 1 of 4 on Nov-2. By Nov-20, the mooring stopped receiving power from two more solar panels (PV2 & PV4) and one of two wind turbines (WT2). This indicates the Power System Controller (PSC) they connect to is failing. Mooring components will be shut down as the mooring loses the ability to recharge its batteries.

Id: 1237 By: crisien
Flag: not_operational Exclude: No

CE07SHSM-MFD37
Method: telemetered
5/25/19, 8:40 AM 10/10/19, 5:07 PM

The high voltage power supply providing power to the seafloor dataloggers and instrumentation failed for reasons unknown at this time. No further telemetered data will be available for this deployment.

Id: 1691 By: wingardc
Flag: not_operational Exclude: No

CE07SHSM-MFD37
Method: recovered-host
5/25/19, 8:40 AM 10/10/19, 5:07 PM

The high voltage power supply providing power to the seafloor dataloggers and instrumentation failed for reasons unknown at this time. No further telemetered data will be available for this deployment.

Id: 1750 By: wingardc
Flag: not_operational Exclude: No

CE07SHSM-MFD37-01-OPTAAD000
10/10/19, 5:30 PM

The OPTAA burst duration was changed from 2 minutes to 4 minutes beginning 2019-10-10T21:30:00 UTC.

Id: 1874 By: wingardc

CE07SHSM-MFD37
Method: telemetered
11/20/19, 1:45 PM 12/4/19, 10:20 PM

Due to reduced power on the mooring (limited recharging of the batteries with light winds and low levels of sunlight), we have had to reduce the operating schedule of the seafloor platform. Normally we run the system hourly, but have reduced that schedule to every 3 hours.

Id: 1833 By: wingardc

CE07SHSM-MFD37
Method: telemetered
12/25/19, 11:45 PM

Power to the seafloor sensors failed. With the exception of the VEL3D and OPTAA, the sensors on the seafloor platform should continue to autonomously sample for the remainder of deployment. Until we recover and redeploy the mooring in the spring of 2020, no further telemetered data will be available from the seafloor sensors.

Id: 1870 By: wingardc
Flag: not_operational Exclude: No

CE07SHSM
Method: telemetered
5/13/20, 6:40 PM

Due to a bug in the mooring telemetry system, data during this time range was not telemetered to shore. It was recorded on the mooring and instruments and will be available in the recovered data streams.

Id: 2185 By: wingardc
Flag: not_available Exclude: No